Hetaida HTD8808E User Manual Download Page 3

Technical File 

File No. 

ZHTF-CE-01-015 

Rev. No. 

A.5

Non Contact Infrared Body 

Thermometer

Instructions for use

Rev. Date 

2020-07-10 

Page 

3 / 19 

3

The  non  contact  infrared  body  thermometer  is  designed  to  minimize  the  possibility  of 
hazards  from  errors  in  the  software  program  by  following  sound  and  light  engineering 
design processes, Risk Analysis and Software Validation. 

Warning 

Warnings are identified by the WARNING symbol shown above. 

  The Non Contact Infrared Body Thermometer is to be operated by consumers in the 

home setting and primary care setting as screening tool. This manual, accessories, 
Directions for Use, all precautionary information, and specifications should be read 
before use. 

  This product is designed to measure human body temperature on the forehead. Do 

not use it for any other purpose. 

  This product is intended in the home setting and primary care setting as screening 

tool.

  Do not use the thermometer if it malfunctions or has been damaged in any matter. 
  When  the  ambient  temperature  of  the  thermometer  changes  too  much,  such  as 

moving the Thermometer from one place of lower temperature to another place of 
higher  temperature,  Allow  the  thermometer  to  remain  in  a  room  for  30  minutes 
where the temperature is between15

to 40

(59

- 104

).

  Remove primary batteries if equipment is not likely to be used for long time. 
  This  product  is  not  waterproof,  do  not  be  immersed  in  water  or  other  liquid;  If 

cleaning  and  disinfection,  please  follow  the  "Care  and  Storage"  section 
requirements.

  Do not touch the sensor of infrared detection with your fingers. 
  If a cold compress on the forehead fever patients, or take other measures to cool 

down the temperature data will low, should be avoided in this case to measure body 
temperature.

  If measure human forehead temperature , please select "body" mode; for measure 

other objects, liquids, food and other temperature please select "surface" mode. 

  This product must be operated in a stable environment, if the ambient environment 

was  mutations,  please  should  be  note  whether  there  is  fog  on  the  sensor,  if  any, 
before using accordance with the "Care and Storage" section to removing the fog. 

  Do  not  near  strong  electrostatic  field  or  strong  magnetic  fields,  thus  avoiding  the 

impact on the accuracy of the measurement data. 

  Do not mix the old and new batteries to avoid damage to the product. 
  It may affect the accuracy of measurements when the forehead is covered by hair, 

perspiration, cap or scarf. 

  The measuring result of this product is only for your reference. If you have any doubt, 

please measure the temperature in other methods. 

Summary of Contents for HTD8808E

Page 1: ...File No ZHTF CE 01 015 Rev No A 5 Non Contact Infrared Body Thermometer Instructions for use Rev Date 2020 07 10 Page 1 24 HeTaiDa Non Contact Infrared Body Thermometer User Manual Model HTD8808E So w...

Page 2: ...and understanding these instructions NOTICE Purchase or possession of this device does not carry any express or implied license to use with replacement parts which would alone or in combination with t...

Page 3: ...a room for 30 minutes where the temperature is between15 to 40 59 104 Remove primary batteries if equipment is not likely to be used for long time This product is not waterproof do not be immersed in...

Page 4: ...fy the device Do not use this device if you think it is damaged or notice anything unusual This device comprises sensitive components and must be treated with caution Observe the storage and operating...

Page 5: ...man body temperature on forehead the skin temperature on one s forehead The operation principle is based on Infrared Sensor technology The IR sensor can output different signal when measuring differen...

Page 6: ...ckaging Refer to operating instructions Compliance with WEEE Standard Operating Temperature DO NOT THROW AWAY Intended for multiple use Operating Humidity Indicates this device is in compliance with M...

Page 7: ...ratory Accuracy Body mode 34 0 34 9 0 3 93 2 F 94 8 F 0 5 F 35 0 42 0 0 2 95 0 F 107 6 F 0 4 F 42 1 43 0 0 3 107 8 F 109 4 F 0 5 F Surface mode 2 3 6 F Display Resolution 0 1 0 1 Three color Backlight...

Page 8: ...m o C m e t I with Equipment classification Safety Standards EN 60601 1 2006 A1 2013 EN 60601 1 2 2015 Type of protection Internally powered equipment on battery power Degree of protection Non Applie...

Page 9: ...e between the eyebrows above and keep the vertical distance press the On Scan button temperature display immediately see figure 2 When the ambient temperature of the thermometer changes too much such...

Page 10: ...e measurement read the data Note 1 After full display over you will hear a rattle or beep beep beep three times which means that the measurements have been completed while the target value of the meas...

Page 11: ...set according to the subjects of different colors and different environments data to meet the different characteristics of populations or individuals HTD8808E as below step 1 Unit Set F1 Under the boo...

Page 12: ...ng software e hardware support Bluetooth 4 0 BLE 3 Troubleshooting MESSAGE SITUATION SOLUTION Temperature taken in not within Typical human temperature range 34 0 43 0 or 93 2 109 4 Make sure the fore...

Page 13: ...g the battery be sure the system is already power off 2 Remove the battery and replace with 2 new one type AAA Make sure align properly as indicated inside the battery cover 4 Slide the battery cover...

Page 14: ...At the end of its life the appliance should not be disposed of in household rubbish Enquire about the options for environment friendly and appropriate disposal Take local regulations into account War...

Page 15: ...nformation provided and this unit can be affected by portable and mobile RF communications equipment 2 Caution Do not use a mobile phone or other devices that emit electromagnetic fields near the unit...

Page 16: ...t 8 KV Air 2 4 8 15 KV Same as the left Electrical fast transient burst IEC 61000 4 4 The input a c power ports 2 KV The input d c power ports 2 KV Signal input output ports 1 KV Not applicable Surge...

Page 17: ...pplies NOTE 2 These guidelines may not apply in all situa ons Electromagne c propaga on is a ected by absorp on and re ec on from structures objects and people a Field strengths from xed transmi ers s...

Page 18: ...mperature runs approximately from 35 5 to 37 8 95 9 F 100 F To determine if one has a fever compare the temperature detected with the person s normal temperature A rise over the reference body tempera...

Page 19: ...nsible for compliance could void the user s authority to operate the equipment This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of th...