Vet/IV 2.2
Infusion Pump
3
Section 3: Warnings, Cautions & Notes
3.1
Warnings, Cautions & Notes Defined
A
WARNING
alerts you about something that could hurt the operator or the patient.
A
CAUTION
alerts you about something that could damage the pump or tubing.
A
NOTE
tells you other important information.
WARNING
This pump is a mechanical device and cannot be guaranteed against failure. For this reason, patients receiving
solutions via this pump should be closely monitored.
WARNING
This pump should be used to deliver intravenous electrolyte and glucose fluids only. Plasma, whole blood or whole
blood products or replacements (
i.e.
, Oxyglobin® Solution) are not recommended for use with the Vet/IV Pump.
WARNING
The Vet/IV Pump is calibrated for either Abbott Series A, Hospira® List No 11545−58 and UniverSet or Abbott Series B
and Baxter ACT 5435 sets. Once calibrated, the pump cannot be used with any alternate system. Please refer to the
label on your pump to ensure that the correct system is being used.
WARNING
This device should not be used in the presence of flammable anesthetics.
WARNING
Do not autoclave, ethylene oxide sterilize, or immerse in liquid. Unplug before cleaning.
WARNING
ELECTRICAL SHOCK HAZARD when covers are removed. Do not remove covers. Service should be performed by
qualified Heska personnel only.
WARNING
Do not use this device in the presence of magnetic resonance imaging (MR or MRI) equipment.
WARNING
Ensure the device’s AC rating is correct for the AC voltage at your installation site before using the pump. The AC
rating is shown on the rear panel rating plate. If the rating is not correct, do not use the pump. Contact Heska’s
Technical Support Services for assistance.
CAUTION
The operator must be thoroughly familiar with the information in this manual before using the pump.
CAUTION
Before operation, verify that administration set tubing is not kinked or occluded.
NOTE:
Operation may be affected in the presence of strong electromagnetic sources, such as electrosurgery equipment,
i.e.,
cautery.
NOTE:
Operation may be affected in the presence of computed tomography (CT) equipment.
NOTE:
Refer to manufacturer’s guidelines for the use of IV sets.