9
15. Open the door by pressing on the door latch. The [door OPEN] alarm should be displayed on the screen.
16. Close the door.
17. [Hold] should appear on the display.
6.4
High Pressure (Occlusion) Alarm Test
18. Occlude the administration set on the delivery (
i.e.
, right) side. Press
RUN/HOLD
to start operation.
19. After a few seconds the [Hi Pres] (high pressure) alarm should activate.
20. Press
RUN/HOLD
to silence the alarm.
21. [Hold] should appear on the display.
22. Release the occlusion in the line.
6.5 Charge Indicator Test
23. Disconnect the power cord from the AC power outlet and verify that the charge indicator turns off.
24. Press
RUN/HOLD
to resume operation. After one second, the display should start flashing to indicate that the pump
is operating on battery power.
25. Connect the power cord to the AC power outlet. Verify that the charge indicator is on and the display has stopped
flashing.