33
Test
frequency
(MHz)
Band
(MHz)
Service
Modulation
Max
Power
(W)
Distance
Test Text
Level (V/m)
Compliance
Level
385
380 – 390
TETRA 400
Pulse 18 Hz
1.8
0.3
27
27
450
430 – 470
GMRS 460, FRS 460
FM ±5 kHz 1
kHz sine
2
0.3
28
28
710
745
780
704 – 787
LTE Band 13, 17
Pulse 217 Hz
0.2
0.3
9
9
810
870
930
800 – 960
GSM 800/900; TETRA 800; iDEN
820; CDMA 850; LTE Band 5
Pulse 18 Hz
2
0.3
28
28
1,720
1,845
1,970
1700 – 1990
GSM 1800; CDMA 1900; GSM
1900; DECT; LTE Band 1, 3, 4,
25; UMTS
Pulse 217 Hz
2
0.3
28
28
2,450
2400 – 2570
Bluetooth; WLAN; 802.11b/g/n;
RFID 2450; LTE Band 7
Pulse 217 Hz
2
0.3
28
28
5,240
5,500
5,785
5100 – 5800
WLAN 802.11a/n
Pulse 217 Hz
0.2
0.3
9
9
17.5. Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment