1.3  Safety provisions

 

The installation of the pole has to be done by a trained person or by one who has set himself 
thoroughly in the assembly instruction. 
 

1.3.1  Warning 
 

 

The product should not be adopted without prior training and before the user reference 
manual has been carefully read and understood. 

 

All kinds of support devices may involve danger if incorrect assembled or wrongly used. 

 

If the pole indicates any sign for loosening, it shall not be used and help be called. 

 

The pole must not be installed in rooms with headroom exceeding the approved or be 
placed where there can be any risk for pinching. 

 

When used in showers friction mat has to be used under the base plate. 

 

If the red warning sign is visible, the assembly is wrong and the pole must not be used. 

 

1.3.2  Installation and assembly 

 
Test before installation 
Beforehand the ceiling should be examined with regard to strength and opportunity for 
placement of the pole at the desired location. At plate-covered ceiling the pole should 
preferably be placed directly under the beams or nailing strip to prevent plate damage. If the 
ceiling is made of concrete, the pole can be placed anywhere. If the wallboard is made of 
gypsum or other week material, The Top plate Big has to be used, or the ceiling has to be 
reinforced by for instance using a more solid wallboard between the pole`s top plate and the 
ceiling. 

 
1.3.3  Installation 

 
The pole is delivered with a loose base- and top plate. 
1.

 

Place the pole on the base plate and set the top plate in place. 

2.

 

Position the pole in a suitable place with the base plate flat against the floor. 

3.

 

Pull out the foot pedal and open the Quick Release clamp. 

4.

 

Rotate the inner tube clockwise so it want slide down by it`s own weight. Extend the pole till 
the top plate reaches the ceiling by pulling the inner tube out of the outer tube. 

5.

 

Close the Quick Release clamp. 

6.

 

Press down the foot pedal. 

7.

 

Check that the red mark is no longer visible. If not, go back to point 4 

8.

 

If the pole is tilted, pull out the foot pedal and adjust it vertical and press down the foot 
pedal again. 

9.

 

Installation of accessories by using the Universal link can now be done. 

10.

 

If the red mark still is visible, this may be the reason: 

 

It is too big distance between the ceiling and the top plate when the rotation of the inner 
tube started. 

 

The quick release could have been too loose. Tighten the nut a few millimeters. 

Summary of Contents for Gripo Alu

Page 1: ...SERVICE MANUAL for Telescopic pole Gripo Alu Telescopic pole Gripo Alu Side 1 av 11...

Page 2: ...products or other queries please contact your supplier or importer Adaptations and Special Adaptations Hepro is a manufacturer and can therefore offer user customized solutions when required Customiz...

Page 3: ...ning 6 1 3 2 Installation and assembly 6 1 3 3 Installation 6 1 3 4 Function Verification and testing of product and accessories 7 1 3 5 Disassembly 7 1 3 6 How to use 7 1 3 7 Error Indications and ch...

Page 4: ...a gas spring with an average pressure of 100 kg This makes the pole possible to be installed without use of any form of screw or nail connection causing perforation of floor or ceiling and the dismant...

Page 5: ...m 40 x 3 0 Average gas spring pressure N 900 Top plate mm 290 Base plate mm 200 Max horizontal loa don the middle of pole kg 100 Product weight without accessories kg 4 7 Max user weight kg 150 MATERI...

Page 6: ...g strip to prevent plate damage If the ceiling is made of concrete the pole can be placed anywhere If the wallboard is made of gypsum or other week material The Top plate Big has to be used or the cei...

Page 7: ...n ring If this shows sign of darkning it should be replaced After this control the pole must be installed in a room where the ceiling is solid and the Quick Release must be checked so that the red mar...

Page 8: ...PRODUCTS WITH SPECIAL LOADS The product has no particular need for periodic inspection if it is a standard product used by a person who in any way exposes the product for special or extreme loads If...

Page 9: ...5100 1 Top plate 1 35110 2 Washer 6 4x20x1 25 3 70277 3 Lock nut Nylock ELZ M6 2 70054 4 Gas spring Stabilus LiftOMat 900N 1 35137 5 Inner tube 1 35111 6 Brand foil red 1 35143 7 Quick release 1 35195...

Page 10: ...SERVICE MANUAL for Telescopic pole Gripo Alu Telescopic pole Gripo Alu Side 10 av 11 18 2 3 1 4 2 3 6 5 7 12 11 8 9 10 13 14 17 15 16...

Page 11: ...ions Edt 02 10 09 2012 Updated 1 3 1 Pos Description No Hepro Item no Universal conection complete 1 35130 20 Hand wheel 1 35123 21 Washer 10 2x40x2 1 70250 22 Bend 40mm 2 35122 23 Adapter bend 40mm 2...

Reviews: