199
Guidance and manufacturer’s declaration- electromagnetic immunity
The model MD300K2 Pulse Oximeter is intended for use in the electromagnetic environment specified
below. The customer or the user of the model MD300K2 Pulse Oximeter should assure that it is used in
such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment-
guidane
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms 150kHz to
80kHz
3V/m 80MHz to
2.5GHz
3V
3V/m
Portable and mobile RF commu-
nications equipment should be
used no closer to any part of the
model MD300K2 Pulse Oximeter,
including cables, than the re-
commended separation distance
calculated from the equation
applicable to the frequency of
the transmitter. Recommended
separation distance
Where P is the maximum output
power rating of the transmitter in
Watts (W) according to the trans-
mitter manufacture and d is the
recommended separation distan-
ce in meters (m). Field strengths
from fixed RF transmitters, as
determined by an electromagnetic
site survey,a should be less than
the compliance level in each fre-
quency range.b Interference may
occur in the vicinity of equipment
marked with the following symbol:
NOTE1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is
Guidance and manufacturer’s declaration- electromagnetic immunity-
For EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration – electromagnetic immunity
The model MD3 0 0 K 2 Pulse Oximeter is intended for use in the electromagnetic
specified below. The customer of the user of the MD3 0 0 K 2 Pulse Oximeter should
assure that it is used in such an environment.
Guidance and manufacturer’s declaration- electromagnetic immunity-
For EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
Immunity test
IEC 60601
test level
Compliance
level
3 Vrms
15 0 kHz
to 8 0 MHz
3 V/m
8 0 MHz
to 2.5 GHz
3 V
3 V/m
Electromagnetic
environment-guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the model MD3 0 0 K 2
Pulse Oximeter, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
Where P is the maximum output
power rating of the transmitter
in Watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
b
transmitters, as determined
by an electromagnetic site
survey,
a
should be less than
the compliance level in each
frequency range.
b
Interference may occur in the
vicinity of equipment marked
with the following symbol:
8 0 MHz to 8 0 0 MHz
8 0 0 MHz to 2.5 GHz
Conducted RF
IEC 6 10 0 0 -4 -6
Radiated
RF IEC
6 10 0 0 -4 -3
d=
3.5
V
1
P
d=
3.5
E
1
P
d=
7
E
1
P
- 29 -
d=1.2 √P
150KHz to 80 MHz
d=1.2 √P
80MHz to 800MHz
d=2.3 √P
800MHz to 2.5GHz