SLOPE ATTN SLOPE
NOTE
This operator’s check checks for proper operation of the
counter chip AlU2, the function and resolution switches,
and the display. This procedure does not check the oper-
ation of the two input amplifiers. See Figure 4-1, Operation
Verification,
for
a more complete operational check.
1.
Set
the line switch
to the ON position.
2. Depress both function switches
check
mode.
@(IN position). This places the HP
5314A
in the self-
3.
Place resolution switch
Hz/N=lOO)
in the IN position. The
HP 5314A
should display
with the overflow
LED
ON and the instrument gating once every second.
4. Place resolution switch
(10
Hz/N=10)
in the IN position. The
HP 5314A
should display
10000.00
with the overflow LED OFF and a 100-millisecond gate time.
5. Place resolution switch
(10
Hz/N=l)
in the
IN
position.
The
HP 5314A
should display
with the overflow LED
and a l-second gate time.
6.
Place both resolution switches
and
in the
IN
position. The
HP 5314A
should display
100000.0
with the overflow LED
OFF and a 100-millisecond gate time.
7.
Place both resolution switches
and
in the
IN
position. The
HP 5314A
should display
100000.
with a IO-millisecond gate time.