SLOPE ATTN SLOPE

NOTE

This operator’s check checks for proper operation of the

counter chip AlU2, the function and resolution switches,
and the display. This procedure does not check the oper-
ation of the two input amplifiers. See Figure  4-1, Operation
Verification, 

for 

a more complete operational check.

1.

Set 

the line switch

to the ON position.

2. Depress both function switches

check 

mode.

 @(IN position). This places the HP 

5314A 

in the self-

3. 

Place resolution switch 

 Hz/N=lOO)

 in the IN position. The 

HP 5314A 

should display

with the overflow 

LED 

 ON and the instrument gating once every second.

4. Place resolution switch 

 (10 

Hz/N=10)

 in the IN position. The 

HP 5314A 

should display

10000.00

with the overflow LED OFF and a  100-millisecond gate time.

5. Place resolution switch

 (10 

Hz/N=l)

 in the 

IN 

position. 

The 

HP 5314A 

should display

with the overflow LED 

 and a l-second gate time.

6. 

Place both resolution switches 

 and 

in the 

IN 

position. The 

HP 5314A 

should display

100000.0

with the overflow LED 

 OFF and a  100-millisecond gate time.

7. 

Place both resolution switches

 and 

in the 

IN 

position. The 

HP 5314A 

should display

100000.

with a IO-millisecond gate time.

Summary of Contents for 5314A

Page 1: ...ow This manual with enclosed Manual Changes sheet applies to HP Model 5314A having serial number prefixes as listed on the Manual Changes sheets Copyright H E W L E T T P A C K A R D C O M P A N Y 197...

Page 2: ...gh this instrument has been designed in accordance with international safety standards this manual contains information cautions and warnings which must be followed to ensure safe operation and to ret...

Page 3: ...h must be followed by the user to ensure safe operation and keep the HP 5314A in safe operating condition 1 8 INSTRUMENTS COVERED BY MANUAL I 9 If the serial number prefix of your HP 5314A is lower th...

Page 4: ...in decade steps of N for N l to 1000 Resolution LSD B Trigger Error X FREQUENCY A N Accuracy 1 count of A B Trigger Error X FREQUENCY A TOTALIZE A Range 10 Hz to 10 MHz Resolution t 1 count of input...

Page 5: ...for the HP 5314A They are a Temperature Compensated Crystal Oscillator TCXO Option 001 and a battery pack for portable operation Option 002 I 17 The seven functions are Frequency Single Shot Period P...

Page 6: ...for use during performance tests adjustments and troubleshooting Substitute test equipment may be used if it meets the required characteristics listed in the table Table 7 3 Recommended Test Equipmen...

Page 7: ...Time Interval Measurements 3 10 The counter measures time intervals from Channel A to Channel B that is Channel A starts the measurement and Channel B stops the measurement Time between points on a s...

Page 8: ...is the same Description numbers match the numbers on the illustration 3 22 OPERATOR S MAINTENANCE 3 23 There is no operator s maintenance for the HP 5314A All maintenance should be per formed by qual...

Page 9: ...2 battery charging circuit in the STBY position Front panel window allowing access to the oscillator frequency adjust ment capacitor AlC2 In HOLD switch IN the measurement except for totalize in progr...

Page 10: ...uses the HP 5314A to make 100 measurements and display the average ratio This switch does not improve accuracy beyond 100 nanoseconds for time interval measurements 0 I 10 Hz N 10 In frequency 10 Hz t...

Page 11: ...n with the attenuator switch 0 to select the relative voltage at which triggering occurs Approximately t350 millivolts is the amount varied The input amplifiers are ac coupled The actual dc level of t...

Page 12: ...ifi 6 Select either 1 Hz or 10 Hz NOTE 10 Hz resolution for frequencies between 10 Hz and 10 MHz may also be used For frequencies higher than 10 MHz the 100MHz 1O Hz switch must be pressed IN NOTE The...

Page 13: ...acy and amplitude 1 Set line switch to the ON position 2 Set COM A SEP switch to SEP position 3 Connect the input signal to INPUT A jack 4 Press PER A T l A B switch IN Make sure the blue shift key is...

Page 14: ...position 3 Press PER A T I A B switch 0 bottom function of switch 0 IN Press the blue shift key IN position This selects the 4 Connect the start time interval signal to the INPUT A jack Connect the st...

Page 15: ...EP position 3 Place the two function switches and in the OUT position This places the HP 5314A in the Ratio A to B function 4 Connect the higher frequency signal to the INPUT A jack D Connect the lowe...

Page 16: ...accuracy and amplitude 1 Set line switch to the ON position 2 Set COM A SEP switch 0 to SEP position 3 Press the FREQ A START A switch IN 4 Press the blue shift key IN position This selects the bottom...

Page 17: ...n the IN position The HP 5314A should display with the overflow LED 0 ON and the instrument gating once every second 4 Place resolution switch 10 Hz N 10 in the IN position The HP 5314A should display...

Reviews: