Model 5314A
Operation
Q
= RATIO A/B I SLOPE ATTN SLOPE
NOTE
See Table 7-7 for the specifications on all input signals
concerning bandwidth, accuracy, and amplitude.
I.
Set line switch
to the ON position.
2. Set COM A/SEP switch
to SEP position.
3. Connect the input signal to INPUT A jack
4. Press FREQ A/START A switch
Be
sure the blue shift
is in the OUT position.
This selects the top function of switch@.
5. Set SLOPE
ATTN
,
cations,
for details.
and LEVEL A
to desired positions; see Table 7-7, Specifi-
6. Select either 1 Hz
10 Hz
NOTE; 10 Hz
resolution
for frequencies between 10 Hz and 10 MHz.
may also be used. For frequencies higher than 10 MHz, the 100MHz/1O Hz
switch
must be pressed IN.
NOTE
The following three resolutions are available with the HP
5314A but are not printed on the front panel.
A. For 0.1 Hz resolution (10 second gate time) on frequenices from 10 Hz to 10 MHz, place all
three resolution switches.
in the OUT position.
B.
For 100 Hz resolution (0.1 second gate time) on frequencies to 100 MHz,
place
switches
and
in the IN position.
c. For 1 kHz resolution (0.01
second gate time) on frequencies to
100 MHz,
place switches
and
in the
IN
position.
Figure 3-3. Frequency Measurement Setup
3-7