39
CALIBRATION OF THE THERMOMETER
Thermometer factory-set. Depending on various skin types and
thickness, there may be temperature differences.
In the switch on state, Press “BODY” button and hold for 3
seconds. the screen will display “F4” , press “SURFACE” button
to increses 0.1°C, press “ROOM” button to reduce 0.1°C, press
“BODY” button to save the setting.
In the cases of seasonal or environmental changes a verification
and adjustment should be carried out.
CHANGING THE BATTERIES
When the LCD screen displays „Battery“ icon,
the batteries are used. Open the lid and
change the batteries, taking great care with
the correct positioning. (“+“ on battery to „+“
in battery compartment and „-“ on battery to
„-“ in battery compartment).
A mistake with this could cause damage of the
device and compromise the quarantee of your thermometer.
Never use rechargeable batteries. Use only batteries for single
usage.
EN