Helmer Scientific i.Series
®
- Platelet Agitator
Operation Manual
360390/A
4
Symbols found on the units
The following symbols may be found on the agitator or agitator packaging:
CE Mark (European units only)
Earth / ground terminal
Caution: Risk of damage to equipment or
danger to operator
Protective earth / ground terminal
Caution: Hot surface
Product falls under the scope of the WEEE
(Waste Electrical and Electronic Equipment)
directive.
Caution: Shock / electrical hazard
1.5
Avoiding Injury
♦
Review safety instructions before installing, using, or maintaining the equipment.
♦
Before moving unit, remove contents from the drawers.
♦
Do not open multiple drawers at the same time.
♦
Before moving unit, disconnect the DC power cord and secure the cord.
♦
When moving unit, use assistance from a second person.
♦
Never physically restrict any moving component.
♦
Avoid removing electrical service panels and access panels unless so instructed.
♦
Keep hands away from pinch points when agitation motion is enabled.
♦
Avoid sharp edges when working inside the electrical compartment.
♦
Ensure biological materials are stored at recommended temperatures determined by standards, literature, or good
laboratory practices.
♦
Proceed with caution when adding and removing products from the platelet agitator.
♦
Use only manufacturer supplied power supply/cord when operating stand-alone or within incubator.
♦ Using the equipment in a manner not specified by Helmer Scientific may impair the protection provided by the equipment.
♦ The platelet agitator is not considered to be a storage cabinet for flammable or hazardous materials.
CAUTION
Decontaminate parts prior to sending for service or repair. Contact Helmer Scientific or your distributor for decontamination
instructions and a Return Authorization Number.
1.6
General Recommendations
General Use
Allow platelet agitator to come to room temperature before switching power on.
During initial startup, motion alarm may sound if the motion is disabled.
During initial startup for stand alone operation, device selection of “X” is required for motion to occur.
Initial Loading
After platelet agitator reaches room temperature, begin storing product.