Pressure connections
“
+, -
”
Hose connectors d = 5 / 6 mm
(0.20 / 0.24 inch)
Housing
PC (polycarbonate)
Fire protection classi
fi
cation UL94V0
Use position
vertical (measuring depends on position)
Protection rating
IP54 according EN 60529
Weight
approx. 230 g
(0.50 lb)
Permissible ambient temperature
-10...60
°
C
(14...140
°
F)
Permissible medium temperature
-10...70
°
C
(14...158
°
F)
Permissible temperature range for storage and
transport
-30...70
°
C
(-22...158
°
F)
Permissible rel. humidity
85 % no condensation
One-sided permissible overload
0.1 bar
(80 in.wg)
Burst prssure
at room temperature: 0.2 bar
(80 in.wg)
at 70
°
C (158
°
F): 0.15 bar
(60 in.wg)
Maximum cross section of terminals
1.5 mm
2
/ AWG16
Interference emission
according EN 61000-6-3 (domestic household applica-
tions)
Interference immunity
according EN 61000-6-2 (industrial applications)
Accuracy and measuring ranges
Pressure measuring range max.
0...1000 Pa
(0...4.0 in.wg)
Tolerance zero point max.
*)
%
+/- 0.9
Tolerance full scale max.
%
+/- 1.3
Resolution
%
0.1
Total of linearity, hysteresis and repeatability max.
%
0.6
Long term stability according to DIN EN 60770
%
+/- 1.0
Temperature coef
fi
cient typical
% / 10K
+/- 0.2
Temperature coef
fi
cient max.
% / 10K
+/- 0.4
Temperature coef
fi
cient sensitivity typical
% / 10K
+/- 0.2
Temperature coef
fi
cient sensitivity max.
% / 10K
+/- 0.4
The accuracy data are percentages and refer to the maximum possible measuring range
of the respective type.
Test conditions: 25
°
C, 45 % RH, voltage supply 12 VDC
*)
Calibration is possible to reduce the zero point error
zero point calibration
Installation and Operating Speci
fi
cation
EDR
Enclosure
L-BAL-E292-GB 1841 Index 001
Part.-No.
12/15