7.2 Read commands
Description
unit according to preselections
1-100% or HI
process runtime
pressure of all connected sensors
operation time in days and hours
Response
XXXX.X mbar/Torr/hPa
XXX%
XX:XX h:m
XXXX.X XXXX.X ...mbar
XXXXdXXh
XXXXXXXXXXXXXXXX
Operation
current
pressure
current speed
time
pressure
operation time
of the controller
device set
preselections
Command
IN_PV_1
IN_PV_2
IN_PV_3
IN_PV_X
IN_PV_T
IN_CFG
0: remote operation off
1: remote operation on
1.......9: sensor quantity
1.......9: sensor number
0: remote module not connected
1: remote module connected
0: level sensor not connected
1: level sensor connected
0: fault indicator not connected
1: fault indicator connected
0: venting valve not connected
1: venting valve connected
0: coolant valve not connected
1: coolant valve connected
0: in-line valve not connected
1: in-line valve connected
0: VMS not connected
1: VMS connected
0: VARIO pump not connected
1: VARIO pump connected
0: acoustic signal off
1: acoustic signal on
0: autostart off
1: autostart on
0: pressure unit mbar
1: pressure unit Torr
2: pressure unit hPa
0......D: language* (hexadecimal)
0: VACUULAN
1: Pump down
2: Vac control
3: Auto mode
4: Program
5: Measuring device
* Language:
0: German
1: English
2: French
3: Italian
4: Spanish
5: Turkish
6: Korean
7: Chinese
8: Portuguese
9: Russian
A: Polish
B: Dutch
C: Japanese
D: Finnish