page 19 of 61
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999248 / 28/04/2011
NOTICE
After switching off the pump, wait 60 sec. minimum until next switching on.
Avoid throttling losses by using connecting pipes with large diameter and keep them
as short as possible.
In case of perturbing exhaust noise connect an exhaust hose.
Install outlet pipelines always falling to avoid backflow of condensate towards the pump.
When assembling, ensure
vacuum-tightness
. After assembly, check the complete
system for leaks.
Secure hose connections at the pump appropriately against accidental detaching.
The controller can only be operated with components which are compatible to the
BUS system, e. g. optional coolant and venting valves.
Pressure transducer of the controller:
Max. permitted pressure at the pressure transducer: 1.5 bar
(absolute).
☞
The display flashes at a pressure higher than 1080 mbar.
Release pressure immediately.
Risk of bursting!
☞
Condensate and deposits in the pressure transducer af-
fect the measuring result.
☞
Clean pressure transducer if necessary.
vacuum
connection
Catchpot:
The round bottom flask is coated with a protective layer to
prevent disintegration in case of breakage or implosion.
➨
Assemble the catchpot using the joint clip.
overpres-
sure safety
relief de-
vice
catchpot at
the outlet
4.3 Condensate cooler (accessory)
Condensate cooler:
➨
Assemble hose nozzles for coolant inlet (1) and coolant
outlet (2) pipelines at the condensate cooler.
The
condensate cooler
enables an efficient condensation of
the pumped vapors at the outlet.
☞
No backflow of condensates.
☞
Controlled recovery of condensates.
☞
Next to 100% solvent recovery.
☞
The isolation cover protects against glass splinters in case
of breaking, acts as thermal isolation to avoid condensa
-
tion of humidity and is intended to absorb shocks.
Attach the pipelines of the coolant circuit to the respective hose nozzles (hose noz-
zles 6-8 mm, see image) at the waste vapor condenser. Check hose connections
prior to starting operation of the cooling system.
Secure coolant hoses at the hose nozzles (e.g. with hose clip) to prevent their ac-
cidental slipping.
outlet (gas!; hose nozzle 10 mm)
NOTICE
(2)
(1)