Display step
(unit: P01)
 mm           inches

Subdi-
vision
P32

Operating Parameter List 

– cont'd.

Parameter

Meaning

Function / Effect

Setting

Counting

Normal 

(

Dir

ection: 

Pos

itive)

direction 

A

Inverse 

(

Dir

ection: 

Neg

ative)

Signal period of encoder 

A

(

Per

iod:) 2, 4, 10, 20, 40, 100, 200

Subdivision of the encoder signals 

A

(

Subd

ivision:) 4, 2, 1, 0.8,  0.5, 0.4, 0.2, 0.1

Select

Error compensation not active

error

Linear error compensation active

compensation

Non-linear error compensation active

Linear error compensation

1)

  

A

Com

pensation – 99 999 < P41 < + 99 999 [µm/m]

Reference

One reference mark

marks 

A

Distance-coded with 500 

 SP

(SP = signal period)

Distance-coded with 1000 

 GP

(e.g. for LS 303 C / LS 603 C)

Distance-coded with 2000 

 SP

Distance-coded with 5000 

 SP

Encoder

Monitoring off 

(

Alar

Off

)

Enc

oder

monitoring 

A

Monitoring on 

(

Alar

On

)

Axis display 

A

Display measured position

(

Axis

)

Do not display measured position /
no encoder

Function of

Resets display to zero

CL key

Does not reset display to zero

1)

Calculate the entry value for P41
Example:

Displayed measuring length La = 620.000 mm

Actual length (determined with, for example, the VM 101 comparator
system from HEIDENHAIN)   Lt = 619.876 mm

Length difference   

L = Lt – La = –124 µm

Compens. factor k: k = 

L / La = –124 µm / 0.62 m = 

–200

 [µm/m]

Parameter Settings for HEIDENHAIN Linear Encoders

LS 303

20

one

single       0.005         0.000 2

4

LS 603

dist.c.

1 000

0.01

0.000 5

2

LB 302

40

one

single

0.01

0.000 5

4

LIDA 10x

dist.c.

2 000

LB 3xx

100

one

single

0.025

0.001

4

dist.c.

1 000

0.05

0.002

2

0.1

0.005

1

Example:

Linear encoder with signal period s = 20 µm
Desired display step  a = 0.005 mm

Subdivision

 P32 = 

0.001 

•••••

 s / a

 = 4

Summary of Contents for ND 510

Page 1: ...point 2 currently active SCL Scaling factor active Define workpiece edge as datum Blinking Waiting for operator to confirm selection Define centerline between two workpiece edges as datum Blinking Wai...

Page 2: ...ay shows the value last assigned to the reference mark position the REF indicator glows and the decimal point blinks Cross over the reference mark in each axis Move the axes one after the other until...

Page 3: ...sed on the new datum This function ends automatically Centerline between two workpiece edges as datum Select the datum Press the SPEC FCT key twice The indicator Centerline as datum starts blinking Pr...

Page 4: ...nput value SPEC FCT Select REF display ENT Save entry Exit compensation value table CL Delete entry Delete all compensation values Enter the parameters and compensation values as follows Display Entry...

Page 5: ...rect input value Input frequency too high for encoder input will occur for example when traverse speed too high Internal counter overflow Error while crossing over reference marks To clear the error m...

Page 6: ...ber Protected Operating Parameters In order to change protected operating parameters the code number 95 148 must be entered via P00 Code They remain accessible until the position display is switched o...

Page 7: ...with 2000 SP Distance coded with 5000 SP Encoder Monitoring off Alarm Off Encoder monitoring A Monitoring on Alarm On Axis display A Display measured position Axis Do not display measured position no...

Page 8: ...le before opening the housing Connect a protective ground This connection should never be interrupted Danger to internal components Do not engage or disengage any connections while the unit is under p...

Reviews: