Working with Scaling Factors

The ND 510 and the ND 550 can display the axis traverse lengthened or shortened by

scaling factor.

 You enter a scaling factor separately for each axis in the user

parameter P12, then activate the scaling factor function with the user parameter
P11. 

SCL

 is highlighted.

Error Messages

Message

Cause and effect

Traverse distance with datum setting function (SPEC FCT) is
too short

Incorrect input value

Input frequency too high for encoder input
(will occur for example when traverse speed too high)

Internal counter overflow

Error while crossing over reference marks

To clear the error message: 

Switch of the display unit.

Should any of these error codes recur, contact your
HEIDENHAIN service agency.

Compensation values for nonlinear axis error compensation have
been erased

Datums have been erased

Erase the operating parameters

If 

all decimal points light up

, the measured value is too large or too small.

 Set a new datum.

To clear error message 

 

           :

When you have removed the cause of the error,

press 

CL

.

Summary of Contents for ND 510

Page 1: ...point 2 currently active SCL Scaling factor active Define workpiece edge as datum Blinking Waiting for operator to confirm selection Define centerline between two workpiece edges as datum Blinking Wai...

Page 2: ...ay shows the value last assigned to the reference mark position the REF indicator glows and the decimal point blinks Cross over the reference mark in each axis Move the axes one after the other until...

Page 3: ...sed on the new datum This function ends automatically Centerline between two workpiece edges as datum Select the datum Press the SPEC FCT key twice The indicator Centerline as datum starts blinking Pr...

Page 4: ...nput value SPEC FCT Select REF display ENT Save entry Exit compensation value table CL Delete entry Delete all compensation values Enter the parameters and compensation values as follows Display Entry...

Page 5: ...rect input value Input frequency too high for encoder input will occur for example when traverse speed too high Internal counter overflow Error while crossing over reference marks To clear the error m...

Page 6: ...ber Protected Operating Parameters In order to change protected operating parameters the code number 95 148 must be entered via P00 Code They remain accessible until the position display is switched o...

Page 7: ...with 2000 SP Distance coded with 5000 SP Encoder Monitoring off Alarm Off Encoder monitoring A Monitoring on Alarm On Axis display A Display measured position Axis Do not display measured position no...

Page 8: ...le before opening the housing Connect a protective ground This connection should never be interrupted Danger to internal components Do not engage or disengage any connections while the unit is under p...

Reviews: