Areas for attachment of the neutral electrode
Attach the neutral electrode to clean, dry skin.
The electrode must be attached close to the operation area, but with a minimum direct distance
from the operating field of 20 cm.
Attach over tissue with a good supply of blood without dips or curves in the skin, for instance on
the upper arm or thigh. (cf. picture below)
Do not attach the neutral electrode to projecting surfaces.
Do not attach above bones, scars, cuts or scratches.
The electrode must not be attached to surfaces with a high amount of hair growth. If necessary,
shave the relevant area before attaching.
Do not attach above particularly adipose areas such as the abdomen or buttocks.
The neutral electrode must not be attached above implants.
Electrodes of monitoring systems must be placed at a distance of at least 20 cm from the operation
area and from the position of the attached neutral electrode
Correct attachment of the neutral electrode in adult patients
Correct attachment of the neutral electrode in children
Attachment in patients with pacemakers
Note! Consult the treating cardiologist!
RIGHT
WRONG!
Summary of Contents for HBS x-touch 200i
Page 6: ...HBS x touch 200i 6 DE Frontansicht des HBS x touch 200i Rückansicht des HBS x touch 200i i ...
Page 28: ...28 ...
Page 31: ...HBS x touch 200i 31 EN ...
Page 34: ...HBS x touch 200i 34 EN Front view of the HBS x touch 200i Rear view of the HBS x touch 200i i ...
Page 56: ...HBS x touch 200i 56 EN ...
Page 58: ...HBS x touch 200i 58 FR ...
Page 84: ...HBS x touch 200 84 FR ...
Page 86: ...HBS x touch 200i 86 ES ...
Page 113: ...113 ...
Page 115: ...HBS x touch 200i 115 IT ...
Page 141: ...PT HBS x touch 200i 141 ...