Specifications
• Product Name: Standard Infrared Ear Thermometer
• Power supply: DC 3V (2 x AAA battery)
• Ear Measuring Range: 82.4°F ~109.4°F (28.0°C ~ 43.0°C)
Ear Measuring Accuracy: ± 0.4°F (or ± 0.2°C)
from 95.9°F ~107.6°F (or 35.5°C ~ 42.0°C) ± 0.5°F ( ± 0.3°C) for other
temperature range
• Object Measuring Range: -4°F ~212°F (-20°C ~ 100°C)
• Object Measuring Accuracy: ± 4°F% or ± 4°F(± 2°C) which ever is greater
• Measurement Time: 1 Second
• Dimension: 1.9 (L) x 1.50 (W) x 6.1 (H) in. [47(L) x 38(W) x 155(H) mm]
• Weight: 83g (with batteries)
• Operation Temperature: 50.0°F ~ 104.0°F (10°C ~ 40°C)
• Operation Relative Humidity: 15% RH – 85%
• Storage & Transportation Temperature: -13°F ~ 131°F (-25°C ~ 55°C)
• Storage & Transportation Relative Humidity: 15% RH ~ 95% RH
• Service Life: 3 years
• Battery Life: Approximately 1 year / 6,000 readings