5
7. OPERATION INSTRUCTION
Master Control Unit Features
2
5
1
ALTERNATE
AUTO
FIRM
STATIC
SOFT
COMFORT
POWER FAILURE
LOW PRESSURE
LOCK
MODE
Health
Co
Royal
Full Automatic System
3
MAX
5
4
8
6
7
Description Item
1. Power Switch – Power On (Green LED), Standby (Amber LED),
2. Pressure Setting Button - allows for manual adjustment of pressures by pressing “Comfort”
button to select a desired comfort from “soft” to “Firm” . Repeatedly pressing the button
will bring the control unit back to “AUTO” mode.
(“AUTO” feature is enabled by default Max)
3. Max – fast inflates the mattress to maximum pressure in static mode to perform nursing
procedure
4. Mode Button-allows for manual select Static or Alternating Mode
(MODE Sequence: DYNAMIC -> STATIC)
Static Mode-press to inflate the mattress to a static surface. The pump will default back to
alternating mode after 20 minutes.
Alternate Mode – inflates mattress in alternating mode
5. Alarm Mute –Press to mute the audio alarm and deferrers to 20 minutes if the problem is
not resolved.
6. Power Failure Indicator- Flashes an Amber LED and sounds an audio alarm to indicate a
power outage situation.
7. Pressure Lost Indicator-Flashes an Amber LED to indicate when control unit fails to inflate
the mattress to the set pressure and will sound 5 mins after the visual alarm.
8. Lock/Unlock button allows user to manually lock/unlock the control panel for unwanted
attempt.
MATTRESS FEATURES
• Alternating therapy to prevent and
treat pressure ulcers
• Modularized design for easy air cell
replacement
• A vapour permeable, pliable and
stretchable nylon top cover provides
low shear, friction and moisture
protection
• Heavy duty polyester bottom cover
Function
Pressure (mmHg)
MAX
60
AUTO
30~50
SOFTER
20
SOFT
25
MEDIUM
30
FIRM
40
FIRMER
50