9
Recommended separation distance between portable and mobile RF communications
equipment and the HealthCo Royal
The HealthCo Royal is intended for use in an electromagnetic environment (for home
healthcare and professional healthcare) in which radiated RF disturbances are controlled. The
customer or the user of the HealthCo Royal can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the HealthCo Royal as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1,2
80 MHz to 800 MHz
d = 1,2
800 MHz to 2,5 GHz
d = 2,3
0,01
N/A
0,12
0,23
0,1
N/A
0,38
0,73
1
N/A
1,2
2,3
10
N/A
3,8
7,3
100
N/A
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1
: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
Manufacturer’s declaration-electromagnetic emissions
The HealthCo Royal is intended for use in the electromagnetic environment (for home healthcare
and professional healthcare) specified below.
The customer or the user of the HealthCo Royal should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment-Guidance
(for home healthcare environment)
RF emissions
CISPR 11
Group 1
The HealthCo Royal uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The HealthCo Royal is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
flicker emissions
IEC 61000-3-3
Compliance