![Health O Meter PRO PLUS 1100KL Operation Manual Download Page 34](http://html1.mh-extra.com/html/health-o-meter/pro-plus-1100kl/pro-plus-1100kl_operation-manual_2113230034.webp)
33
Name
ESC
character
ESC value
with
parameters
Description
Response
P P
ccc=ccc
cccc
This will be the scales response to the request mentioned above
Diagnostics(1)
A A
ccc
This is the request for a diagnostic test on certain parts of the scale
(such as battery life, loadcells, keypad, LCD etc.). (See
Table B
)
Diagnostics(2)
Z Z
ccc
This will be the response of the diagnostics done on the scale.
Values will include error codes to indicate what is wrong with the
scale, or all zeros to indicate that all is well.
Control (set a
value)
C C
ccc=ccc
ccc
his is to set the value of the scale's global settings such as backlight,
auto hold time, keypad beep volume etc. (See
Table A
) (i.e.
<ESC>CAHT=20<ESC>E will set the Auto Hold Time to 20 seconds)
Get subject
record data and
readings
G G
This will return all the information from both the subject record data
as well as the readings stored in the readings table.
Set subject
ecord data
S S
This will set the subject record data for a specific patient.
Immediately following this will be the value that is set. (i.e.
<ESC>S<ESC>I1234<ESC>T200<ESC>Um<ESC>E will set the
TARE for patient ID1234 to 200kg)
Clear readings
X X
This will clear all the values stored in the readings table on the scale.
Only the subject record data will not be deleted.
Clear subject
record data
Y Y
This will clear the subject record data for a specific patient. The
patient ID will follow the Y immediately (<ESC>Iccccccccccc). If no
ID is given, ALL the data will be cleared.
Name of Control
Identifier Unit
Auto Hold Time
AHT
nn (Seconds)
Auto Off Time
AOT
nn (Minutes)
Sleep Time
STM
nn (Seconds)
Keypad Volume
VKE
nn (01..10)
Stable weight Volume
VSW
nn (01..10)
Alarm Volume
VAL
nn (01..10)
Display of Date/Time
DAT
nn (01 or 10)
Backlight BAC
nn
(01..10)
Contrast CON
nn
(01..10)
Weight Display Mode (fractions(f), decimals(d)
or Pounds and ounces (p)
WDM
c (f,d,p)
Tare Weight
TAR
nnn.nn (Kg)
Unit Of Measure (Metric or constitutional)
UOM
c (m or c)
Positions left in RAM
RAM
nnnn (0..
max
)
Table A - Scale Global Values List and Identifiers
Name of component to test
Identifier
Error Code
Load Cell A
LCA
E3A
Load Cell B
LCB
E3B
Load Cell C
LCC
E3C
Load Cell D
LCD
E3D
Battery
BAT
E4L(Bat
L
ow, but still usable) or
E4U (Bat Low and
U
nstable)
PC Communication (USB)
PCC
E05
Write to RAM
WRM
E08
Read from RAM
RRM
E09
Table B - Components to test
Summary of Contents for PRO PLUS 1100KL
Page 29: ...28 MODEL 2650KL EXPLODED VIEW OF SCALE...
Page 31: ...30 MODEL 2700KL EXPLODED VIEW OF SCALE View of Scale in Folded Position...
Page 68: ...67 MODELE 2650KL SCHEMA DE LA BALANCE EN PIECES DETACHEES...
Page 70: ...69 MODELE 2700KL SCHEMA DE LA BALANCE EN PIECES DETACHEES...
Page 106: ...105 MODEL0 2650KL VISTA DETALLADA DE LA BALANZA...
Page 108: ...107 MODELO 2700KL VISTA DETALLADA DE LA BALANZA Vista de la Balanza en Posici n Plegada...