24
CALIBRATION PATH
Calibration is performed using kg or lb, according to the units of measure used upon
entering into the calibration path.
01-Cal With One Point process below:
Operator Action
Display
1. Make sure the scale is turned off, then press
and hold the HOLD/RELEASE button while you
press the On/OFF button.
Enter load weight:
450.0LB( or 200.0KG)
2
ˊ
Input the weight , then press ENTER button.
Enter load weight:
450.0LB( or 200.0KG)
3. Make sure there is nothing on the platform, then
press ENTER button.
Zero calibration
Please clear the scale
4. Wait for about 1-3 second, Zero calibration
finished.
Zero calibration
Please wait
5. Put the weight on the platform, then press
ENTER button.
Weight calibration
Put: XXX.X
6. It will calibrate after 1-3 seconds.
Weight calibration
Please wait
7. If success, display shows Factor OK!. If failed,
display Factor Fail.
Factor Update
Factor OK!
˄
or Factor Fail!
˅
8. If successful the process will clear the weight.
Rebooting…
Please wait
9. The scale will reboot after wait for about 1-2
seconds.
Health o Meter
Pro-Plus
10. Please wait until the scale resumes normal
operation. Remove the weight from the platform.
Health o Meter
Pro-Plus
Your Pro Plus scale has been calibrated at the factory and does not require
re-calibration prior to use.
Summary of Contents for PRO PLUS 1100KL
Page 29: ...28 MODEL 2650KL EXPLODED VIEW OF SCALE...
Page 31: ...30 MODEL 2700KL EXPLODED VIEW OF SCALE View of Scale in Folded Position...
Page 68: ...67 MODELE 2650KL SCHEMA DE LA BALANCE EN PIECES DETACHEES...
Page 70: ...69 MODELE 2700KL SCHEMA DE LA BALANCE EN PIECES DETACHEES...
Page 106: ...105 MODEL0 2650KL VISTA DETALLADA DE LA BALANZA...
Page 108: ...107 MODELO 2700KL VISTA DETALLADA DE LA BALANZA Vista de la Balanza en Posici n Plegada...