21
TEDS
A1813-1.1 en
HBM
HBM template: User Channel Comment
Parameters
Value
(example)
Change requires
rights to level:
Explanation
User Channel Comment
my transducer
USR
Comment of up to 45
characters in 7-bit ASCII
IEEE template
Calibration Curve:
Parameters
Value
(example)
Unit
Chang
e re-
quires
rights
to
level:
Explanation
Domain parameter of the Cal-
ibration Curve
Physical
CAL
Calibration curve segments
1
Elements
CAL
1. Starting domain value of
segment (% of full span)
0
%
CAL
Calibration curve polynomial
3
Elements
CAL
1. Power of domain value
1)
1.0
CAL
y = a
1
x
1
+ a
2
x
2
+ a
3
x
3
Power for the first poly-
nomial element (1 in the
example)
1. Polynomial coefficient
9.999E+001
CAL
y =
a
0
x
0
+ a
1
x
1
2. Power of domain value
2.0
CAL
2. Polynomial coefficient
2.000E-002
CAL
3. Power of domain value
3.0
CAL
3. Polynomial coefficient
-5.000E-003
CAL
NOTE
Range is always from 0% to 100%; i.e. when the lowest value, e.g.
is -200 N, this corresponds to 0% (and not -100%).
1)
a
0
x
0
= measured value;
a
0
= Polynomial coefficient