29
TEDS
A1813-1.1 en
HBM
•
Select the card file ”
Characteristic curve
”, enter the nominal measuring
range of the transducer and the respective sensitivity (from the test
certificate or calibration certificate).
With the nominal measuring range, it must be noted that values must be
entered with the Pascal unit (1 bar = 100000 Pa), in this case
10 MPa.
NOTE:
The general unit bar is not supported because of the Standard IEEE
1451.4. However, to display the unit bar, the additional HBM specific
Template Unit Conversion is used. This is available by default and set up
for Pascal/bar conversion in the templates for TEDS contents that are
supplied since the version
1.0.1
of the TEDS Editor / version
1.2.7 of the
MGCplus Assistant. Furthermore, the firmware used in the amplifier should
at least comply with the version specified belows.
Single-channel modules (MLxxB)
P5.50
ML455
P6.28
ML460
P1.24
ML801B
P6.26