17
TEDS
A1813-1.1 en
HBM
Parameters
Value
1)
Unit
Chang
e re-
quires
rights
to
level:
Explanation
Input Impedance @
nominal frequency
102
Ohms
ID
Input resistance according to the
HBM data sheet.
Calibration Date
10-Dec-2003
CAL
Date of the last calibration or cre-
ation of the test certificate (if no cal-
ibration carried out), or date of stor-
ing the TEDS data (if only nominal
values from the data sheet were
used).
Format: day-month-year.
Abbreviations for the months: Jan,
Feb, Mar, Apr, May, Jun, Jul, Aug,
Sep, Oct, Nov, Dec.
Calibration Initials
HBM
CAL
Initials of the calibrator or calibration
laboratory concerned.
Calibration Period
(Days)
730
days
CAL
Time before recalibration
Measurement location
ID
0
USR
Identification number for the mea-
suring point. Can be assigned ac-
cording to the application. Possible
values: a number from 0 to 2047. If
that is not enough, the HBM Chan-
nel Comment template can also be
used for this purpose.
1)
Typical values for a type WA displacement transducer, measuring range 50 mm, from HBM