Template: Frequency/Pulse Sensor
Parameter
Value
Unit
Require
d user
rights
Explanation
Transducer Electrical
Signal Type
Pulse
Sensor
ID
Minimum Torque
0.000
N
@
m
CAL
The physical measurand and
unit are defined when the
template is created, after which
they cannot be changed.
Maximum Torque
1000
N
@
m
CAL
Pulse Measurement Type
Frequency
Minimum Electrical
Value
10000
Hz
CAL
The difference between these
values is the nominal (rated)
sensitivity.
Maximum Electrical Value
15000
Hz
CAL
Mapping Method
Linear
Discrete Signal Type
Bipolar
ID
Discrete Signal Amplitude
4
V
Discrete Signal
Configuration
Single
Transducer Response
Time
0
secon
ds
Excitation Level nom
24
V
Excitation Level min
18
V
Excitation Level max
30
V
Excitation Type
DC
Excitation Current draw
0.5
A
Calibration Date
1-Nov-2006
CAL
Date of the last calibration or
creation of the manufacturing
certificate (if no calibration
carried out), or of the storage of
the TEDS data (if only nominal
(rated) values from the data
sheet were used).
Format: day-month-year.
Abbreviations for the months:
Jan, Feb, Mar, Apr, May, Jun,
Jul, Aug, Sep, Oct, Nov, Dec.
Calibration Initials
HBM or PTB
CAL
Initials of the calibrator or
calibration laboratory
concerned.
Calibration Period
(Days)
0
days
CAL
Time before recalibration,
calculated from the date
specified under Calibration
Date.
Measurement location ID
0
USR
Identification number for the
measuring point.
Can be assigned according to
the application. Possible
values: a number from 0 to
2047.