![HBK FS22SI User Manual Download Page 80](http://html1.mh-extra.com/html/hbk/fs22si/fs22si_user-manual_2111679080.webp)
FS22SI
BRAGGMONITOR SI USER INTERFACE
80
Fig. 7.29
Zoomed spectum
Saving Spectral Data
To save spectral data, press the “save data" button (number
8
in
), then select the
file path to define the folder where data is stored and define the file name. Data sets will
be recorded in .txt format for a single sweep (
matically return to its default state.
Saved data file is organized in columns (
), meaning:
1
1
st
column – wavelength values in nanometers
2
2
nd
column – optical power values in dBm.
Fig. 7.30
Saved spectrum file
Summary of Contents for FS22SI
Page 1: ...FS22SI Industrial BraggMETER SI ENGLISH User Manual...
Page 18: ...FS22SI OPERATION 18 Fig 3 7 Local Area connection properties...
Page 61: ...61 FS22SI BRAGGMONITOR SI SOFTWARE DETAILS Fig 6 2 Set Run as administrator automatically...
Page 88: ...FS22SI BRAGGMONITOR SI USER INTERFACE 88 Fig 7 40 Update info window...
Page 89: ...89 FS22SI BRAGGMONITOR SI USER INTERFACE...
Page 90: ...HBK Hottinger Br el Kjaer www hbkworld com info hbkworld com A04249 10 E00 00...