13
FS22SI
OPERATION
edge proper power supply, the POWER LED will turn green during 2 seconds after con
necting the power supply.
Information
For the Standard version of the interrogator the power supply must have a minimum of
30 W, with an output voltage between 12-36 v and must withstand a peak of 4 A for at least
4 ms. For the Rack-Mountable version of the interrogator the power must be 100-240 V AC,
50-60 Hz, 1A.
Notice
Powering above the specified limits will damage the equipment. For the Rack-mountable
version there is a fuse protection that can be replaced. please see section 4.4.5 Fuse fail
ure.
Information
The interrogator must be powered using a power supply source not shared with other
equipment's. A replacement AC-DC power supply unit 24 V (1-NTX001) for the Standard
version or Power Cable (1-KAB274) available with different plug formats can be ordered
from HBK.
3.2.2
Optical Connectors
The FS22 - Industrial BraggMETER SI can be purchased either with FC/APC, or SC/APC
optical connectors.
number
1
exemplifies FC/APC connectors.
Attention should be paid to the cleaning of the optical connectors. A dirty connector can
compromise the measurement and will degrade the interrogator performance. It is advis
able to frequently clean the connectors using appropriate tools.
FS22 - Industrial BraggMETER SI can have one, four or eight optical channels in parallel.
3.2.3
Ethernet Connection
Connect the Ethernet RJ45 connector directly to a PC using a Ethernet cross-over cable,
or to a network connector using a direct Ethernet cable (in this case, the FS22 - Industrial
BraggMETER SI must be in the same subnet as your Local Network).
Important
The interrogator default network configuration is “10.0.0.10:255.0.0.0:0.0.0.0".
Summary of Contents for FS22SI
Page 1: ...FS22SI Industrial BraggMETER SI ENGLISH User Manual...
Page 18: ...FS22SI OPERATION 18 Fig 3 7 Local Area connection properties...
Page 61: ...61 FS22SI BRAGGMONITOR SI SOFTWARE DETAILS Fig 6 2 Set Run as administrator automatically...
Page 88: ...FS22SI BRAGGMONITOR SI USER INTERFACE 88 Fig 7 40 Update info window...
Page 89: ...89 FS22SI BRAGGMONITOR SI USER INTERFACE...
Page 90: ...HBK Hottinger Br el Kjaer www hbkworld com info hbkworld com A04249 10 E00 00...