![HBK FS22DI User Manual Download Page 64](http://html1.mh-extra.com/html/hbk/fs22di/fs22di_user-manual_2111693064.webp)
FS22DI
BRAGGMONITOR DI USER INTERFACE
64
Formula
Function defining the correlation between the wavelength
shift (Δλ=x) and the engineering values.
x corresponds to the difference between the measured wave
length (λ) at each instant and the defined central wavelength
(λ
0
).
Here it is possible to set a formula manually, although it is also
possible to copy and paste from other locations. If the desired
output is wavelength shift, insert x on formula textbox.
Do
Do Not
-11.3*x
-11.3x
-11.3*x^2+105.4*x+30
=-11.3*x^2+105.4*x+30
Note: For decimal separation, use comma or dot according to
the regional settings of the computer.
Tab. 7.1
Save and Load configuration
Acquisition configuration can be saved in .txt extension file by pressing the “save” button
(number
9
in
) and choosing a save path as shown in
.
Fig. 7.21
Save configuration