BRAGGMONITOR DI USER INTERFACE
Fig. 7.20
Sensor Settings
Sensor Name
Sensor's identification field. It is not possible to have two sen
sors with the same name.
Central Wavelength
(CWL)
Reference wavelength from which wavelength shift (Δλ=x) is
calculated.
Range
Safety bands defined to avoid crosstalk between sensors.
Each sensor has its own operation range. For instance, if we
have a central wavelength of 1522.659 nm and a range span
ning 3 nm, only wavelength values between 1525.659 and
1519.659 are accepted. Wavelength values out of the defined
bands are returned as -998.