HAUBER-Elektronik GmbH
13
13.3
Operating range of the vibration monitoring unit
The operating range is independent from the measuring range. It can be derived from the
maximum acceleration, which is 16.5 g across all frequencies. The maximum measurable
vibration velocity is based on the formula
𝑣
𝑚𝑎𝑥
= ∫ 𝑎
𝑚𝑎𝑥
For sinusoidal vibration,
𝑣
𝑚𝑎𝑥
=
𝑎
𝑚𝑎𝑥
2𝜋𝑓
applies
Fig. 1: shows the operating range of the vibration monitoring unit, which is limited by the
maximum measurable vibration velocity in mm/s depending on the frequency in Hz.
Fig. 1: Operating range diagram
1
Frequency in Hz
2
Vibration velocity in mm/s
3
Operating range of the vibration monitoring unit
Reading example:
Frequency
(Hz)
Maximum measurable
Vibration velocity
(mm/s)
250
103
400
64
1000
25
Tab. 3: Reading example of operating range
0
20
40
60
80
100
120
140
160
180
200
0
100
200
300
400
500
600
700
800
900
1000
2
→
1
→
3