Test Tube and Heated Cell Holder Installation
1.
Lay the instrument upside down on a soft surface.
2.
Unscrew the two screws and remove the cuvette cell holder by tilting the holder to allow the cell ring to clear the
baseplate.
3.
There are two positioning areas to help indicate the orientation of the cell holders when installing. The cell holders
will not fit unless this is correctly aligned.
4.
Insert the new cell holder and screw in the two screws.
Align towards the front
of the instrument
Align towards the back
of instrument
Front of instrument
Back of instrument
Front of instrument
Back of instrument