26 |
P a g e
A graph will display the results and the fitted curve as
the measurements are input. Use the up and down
arrows to select a standard to be repeated if a poor
reading has been obtained. Use C to clear the
previous reading.
Step 17
Press
to accept the calibration and go to the
Results screen or press Back
to return to the
Standards screen.
Calibration (Manual and New Standards)
This shows the calibration values and allows
standards to be measured.
Step 18
Enter the concentration values by using the keypad
numbers and the up and down arrows to move
between the different standard boxes. Range 0.001
to 9999.
Press the down arrow key.
Repeat Step 18 for all standards
Step 19
Press Next
when all standards have been
entered. If any duplicate or non-monotonic
(increasing entries) are present the unit will beep and
highlight the incorrect entry or press back
to
return to the Parameter screen.
Step 20
Select whether or not to adjust the name of your
sample before each new run, using left and arrow
keys to select Yes or No. Press
.