21
English
EN
Error message Possible causes
Remedy
Implausible
measured
values
Implausible measured
values often occur
due to inappropriate
handling of the device
or if mistakes were
made during measuring.
Please follow the directions
for correct use given in the
instructions for use and all
important information given in
Chapters 3, 7 and 8.
Then repeat the temperature
measurement.
If body temperature values
remain implausible, please
contact your doctor!
■
Switch the device off if an error message appears.
■
Check for possible causes and follow the directions for correct use in the
instructions for use, and in particular the instructions regarding measurement
in Chapter 3 „Important information“ and in Chapters 7 and 8.
■
Wait for 1 minute and then repeat the temperature measurement.
11. Cleaning and care of the device
The measuring sensor is the most important and most sensitive part of
the device. To ensure accurate measurement, the sensor must be clean and
undamaged at all times.
■
Please do not use any aggressive cleaning agents or solvents as these can
damage the device and obscure the screen.
■
The device is not waterproof. Make sure no liquid can penetrate the inside
of the thermometer and never immerse the device in water or other
cleaning agents!
■
Clean the device and the measuring sensor with a moist cloth and a mild
detergent or a suitable disinfectant (e.g. 70% isopropanol).
■
After cleaning the measuring sensor, you need to wait until the cleaning
liquid has dried off completely before taking a temperature measurement.
■
To protect from external influences, you can store the thermometer in the
storage box.