4. Current supply

Open the battery cover on the
underside of the device. Insert 
batteries (see chap. 12, Technical
specifications), ensuring correct
polarity (“+” and “–”). Close the
battery lid.

If you do not intend to use the 
device for a longer period of 
time, please remove the batteries.

In the interests of environmental 
protection, used batteries should 
not be disposed of in household 
waste. Please observe current 
regulations when disposing of 
the batteries or use municipal
collecting bins.

Do not discard with household 
waste, but use specially 
designated disposal facilities 
(recycling centres etc.).

flashing arrow pointing down-
wards shows that the device is 
checking itself automatically. A 
beep sounds and the “0” appears
in the display which signals that 
the unit is now ready to take a 
new measurement. If the down-
ward arrow symbol “   ” is dis-
played, there is some air remain-
ing in the cuff. Release air using 
the exhaust button until the “0” 
appears.

Pressurize the cuff using the 
pumping bulb until you have a 
pressure value which is suitable 
for you. It should be approxi-
mately 30 mmHg higher than 
your regular systolic blood pres-
sure and is shown on the display.
When pressurization is com-
pleted, Tensoval compact 
automatically begins the meas-
urement. If the cuff was not 
inflated enough or if the meas-
urement is disturbed, the arrow 
pointing upwards “   ” appears. 
Pump again until a high enough 
pressure is reached.

Important: do not move or talk 
throughout the entire measur-
ing procedure.

As pressure in the cuff decreases, 

cuff is approx.

2.5 cm away from

the bend of the elbow

(see fig.).

The cuff should be firm but not 
too tight. You should be able to 
push one or two fingers between 
the arm and the cuff. Note: 
Uneven wrapping of the cuff can 
lead to inaccurate results.

7. Measuring blood pressure

It is recommended that blood 
pressure measurements are car-
ried out whilst seated. Place your
arm with the palm upwards on a 
support and make sure the 

cuff 

is at the level of the heart

Do not turn the device on until 
the cuff has been applied, other-
wise the cuff can become dam-
aged through the resulting 
excess pressure.

Press the

START/STOP button

. All 

display symbols appear followed 
by the last measured value. A 

Pumping bulb tube

Cuff tube

English

ca. 2.5 cm

11

10

5. Connecting the tubes

6. Applying the cuff

Measurement should be taken 
on 

the naked

arm which usually 

has the higher blood pressure 
value.

When using the 

draw-clamp 

cuff

, thread the end of the cuff 

through the metal hoop if neces-
sary, so as to form a loop. In this 
instance the Velcro fastener must
be on the outside. Wrap the cuff 
round the upper arm, the tube 
lies in the centre of the elbow 
bend, pointing towards the 
hand. Take the free end of the 
cuff, wrap it firmly round the arm
and close the Velcro fastener.

Check that the lower edge of the 

All manuals and user guides at all-guides.com

all-guides.com

Summary of Contents for Tensoval compact

Page 1: ...ovalR compact Instructions for use Instrukcja obs ugi Haszn lati utas t s N vod na pou vanie Instrucciones de uso Manual de utiliza o All manuals and user guides at all guides com a l l g u i d e s c...

Page 2: ...addresses are given in the warranty document On the Internet you can visit our homepage on www hartmann info Prosimy dok adnie zapozna si z niniejsz instrukcj obs ugi poniewa prawid owy pomiar ci nie...

Page 3: ...stechnol gi val lehet v teszi gyorsan s megb zhat an a systol s s diastol s v rnyom s m r s t valamint a pulzussz m oszcillom teres m r si elj r ssal t rt n meghat roz s t HARTMANN Tensoval compact S...

Page 4: ...imal up to 120 mmHg up to 80 mmHg Normal up to 130 mmHg up to 85 mmHg Normal limit values 130 139 mmHg 85 89 mmHg Grade 1 hypertension 140 159 mmHg 90 99 mmHg Grade 2 hypertension 160 179 mmHg 100 109...

Page 5: ...conditions it leads to are at an advanced stage These may be life threatening such as heart attack stroke or kidney failure Only regular blood pressure monitoring enables high blood pressure to be det...

Page 6: ...g upwards appears Pump again until a high enough pressure is reached Important do not move or talk throughout the entire measur ing procedure As pressure in the cuff decreases cuff is approx 2 5 cm aw...

Page 7: ...document the heart symbol and the falling cuff pressure are displayed A beep sounds which is synchron ised with your heartbeat A beep indicates the end of measurement The systolic and diastolic readin...

Page 8: ...Cuff was applied incorrectly or too loosely Measurement was disturbed due to speaking or moving The pulse could not be measured accurately The battery is empty Replace the battery The device has a te...

Page 9: ...uring systems EN 1060 3 1997 Clinical testing of measurement accuracy was performed according to the testing protocol recommended by EN 1060 3 according to DIN 58130 and ANSI AAMI SP10 1992 Manufactur...

Page 10: ...ymalne do 120 mm Hg do 80 mm Hg Normalne do 130 mm Hg do 85 mm Hg Warto graniczna normalna 130 139 mmHg 85 89 mmHg Hipertonia 1 stopnia 140 159 mmHg 90 99 mmHg Hipertonia 2 stopnia 160 179 mmHg 100 10...

Page 11: ...ci nienia t tniczego jest niezwykle wa na w okresie ci y gdy ci a prowadzi mo e do zmiany tego ci nienia Jednak wyniki pomiar w nale y interpre towa wy cznie w porozumieniu z lekarzem 3 Symbole na wy...

Page 12: ...arat w cza dopiero po na o eniu mankietu gdy w przeciw nym razie mankiet ulec mo e zniszczeniu wskutek zbyt du ego wytwarzaj cego si w nim ci nienia Nacisn przycisk START STOP Na wy wietlaczu uka si w...

Page 13: ...alecamy przeprowadzanie co 2 lata kontroli technicznej w zakresie sprawno ci pomiarowej aparatu Kontrola techniczna pomiaru mo e by przeprowadzona przez odpo wiednie s u by lub autoryzowany serwis 10...

Page 14: ...ez minut i pomiar wykona ponownie Podczas pomiaru nie wolno wykonywa ruch w ani m wi 27 Polski 12 Dane techniczne Metoda pomiaru oscylometryczna Zakres wskaza 0 280 mmHg Zakres pomiaru ci n skurcz SYS...

Page 15: ...nia krwi EN 1060 3 1997 Kliniczn kontrol dok adno ci pomiaru przeprowadzo no wed ug protoko w kontrolnych zalecanych przez EN 1060 3 wed ug DIN 58130 oraz ANSI AAMI SP10 1992 Producent PAUL HARTMANN A...

Page 16: ...r az art ri s rrendszerbe s ell tja a testet oxig nnel s a t bbi sz ks ges t panyaggal Ahhoz hogy a v r raml s a legkisebb v red nybe is eljusson sz ks g van egy lland nyom sra a v rnyom sra A v r ny...

Page 17: ...ek nt egyes esetekben helytelen m r si eredm nyek sz lethetnek ill nem kapunk m r si eredm nyt Err A v rnyom s ellen rz se a ter hess g ideje alatt rendk v l fon tos mert a v rnyom s a terhes s g k ve...

Page 18: ...r m r sre k sz llapotban van Amennyiben a kijelz n a lefel mutat ny l megmarad az azt jelzi hogy a rendszerben m g leveg tal l hat A leereszt gomb seg ts g vel engedje ki az sszes leve g t m g a kijel...

Page 19: ...A k sz l ket kiz r lag puha megnedves tett ruhadarabbal szabad tiszt tani Tiszt t vagy old szert nem szabad felhaszn lni A mandzsett t vatosan kicsit megnedves tett ruhadarabbal s gyenge szappanoldat...

Page 20: ...kt roz si sz ll t si k vetelm nyek k rnyezeth m rs klet 20 C 60 C relat v l gnedvess g 10 95 Sz riasz m az elemtart ban Fell pett hib k Lehets ges okok Nem rt kelhet m r si Helytelen mandzsetta m ret...

Page 21: ...ek az elektromechanikus v rnyom sm r rendszerek r sz re A m r spontos s g klinikai vizsg lata az EN 1060 3 szerint aj nlott vizsg lati protokoll alapj n lett elv gezve DIN 58130 s ANSI AAMI SP10 1992...

Page 22: ...1 60 80 120 2 5 42 43 P 120 80 130 85 130 139 85 89 140 159 90 99 160 179 100 109 180 110 All manuals and user guides at all guides com...

Page 23: ...11 B 4 12 45 P 22 32 Err 3 44 All manuals and user guides at all guides com...

Page 24: ...2 5 7 START STOP 0 0 30 Tensoval compact 47 P 5 6 46 2 5 All manuals and user guides at all guides com...

Page 25: ...2 10 49 P START STOP 3 8 9 48 11 280 All manuals and user guides at all guides com...

Page 26: ...TMANN 2 1 12 0 280 SYS 50 270 DIA 30 250 40 200 3 5 EN 1060 3 4 x 1 5 Mignon R6 LR6 3 22 32 1 10 C 40 C 30 85 20 C 60 C 10 95 50 51 P All manuals and user guides at all guides com a l l g u i d e s c...

Page 27: ...compact 93 42 CE EN 1060 1 EN 1060 1 1995 3 EN 1060 3 1997 EN 1060 3 DIN 58130 ANSI AAMI SP10 1992 PAUL HARTMANN AG 89522 Heidenheim 14 HARTMANN 900 179 22 32 900 190 52 All manuals and user guides a...

Page 28: ...h faktorov napr d chanie zhlboka u vanie povzbudzuj cich prostriedkov rozpr vanie rozru enie klima tick faktory m u vies k fluktu ciam krvn ho tlaku V d sledku toho v doktor a lek rnik sa asto dopracu...

Page 29: ...lneho odpadu Nelikvidujte s domov m odpa dom ale na peci lnych miestach k tomu ur en ch napr centr zberu odpadu T mto sp sobom pom ete ivotn mu prostrediu 5 Napojenie hadi iek Nasu te hadi ku pumpi k...

Page 30: ...slu ky V tejto polohe mus by plocha such ho zipsu na vonkaj ej strane Obto te man etu okolo ramena tak aby hadi ka le ala v strede lak ovej jamky a smerovala ku dlani Uchopte voln koniec man ety pevne...

Page 31: ...ne lek rske praxe a nemocnice Kontrola funkcie merania m e by za poplatok vykonan bu v autorizovan ch organiz ci ch alebo u kvalifikova n ch oprav rov 10 Z ruka Na tento pr stroj sa vz ahuje pred en z...

Page 32: ...o vajte a potom meranie zopakujte Po as merania sa neh bte a nerozpr vajte 12 Technick parametre Met da merania oscilometrick Rozmedzie displeja 0 280 mmHg Rozmedzie v meran Systola SYS 50 270 mmHg Di...

Page 33: ...ie krvn ho tlaku pod a EN 1060 3 1997 Klinick testy presnosti merania boli vykonan v zhode s testovac m protokolom odporu en m v norme EN 1060 3 v zhode s DIN 58130 a ANSI AAMI SP10 1992 V robca PAUL...

Page 34: ...e y a lo largo de un extenso per odo de tiempo Con la ayuda de los valores de tensi n arterial que usted haya registrado cada d a regularmente la hiperten si n puede detectarse muy pronto y su doctor...

Page 35: ...ectuadas con este dispositivo pueden resultar falsas o bien no llegar a realizarse en algunos casos Err La monitorizaci n de la tensi n arterial es extraordinariamente importante durante la gestaci n...

Page 36: ...az n No active el dispositivo hasta despu s de colocar el brazalete si no act a de este modo el bra zalete puede resultar da ado debido al exceso de presi n Presione el bot n de INICIO PARADA Todos lo...

Page 37: ...ado bastante Bombee de nuevo hasta alcanzar una presi n lo bastante alta El brazalete se ha inflado a m s de 280 mmHg Desinfle el brazalete pulsando el bot n de escape El brazalete se ha aplicado inco...

Page 38: ...automedici n en el Cap tulo 2 Rel jese durante un minuto y vuelva a tomar la medici n No se mueva ni hable durante la misma 12 Especificaciones t cnicas M todo de medici n oscilom trico Intervalo de p...

Page 39: ...yos cl ni cos de precisi n de la medici n se realizaron de conformidad con el protocolo de ensayo recomendado por la EN 1060 3 de conformidad con la DIN 58130 y ANSI AAMI SP10 1992 Fabricante PAUL HAR...

Page 40: ...s surgem num estado avan ado das doen as consequentes Estas no entanto podem ser mortais como o enfarte do mioc rdio o acidente vascular cerebral ou a insufici ncia renal Uma tens o arterial elevada s...

Page 41: ...aparelho Introduza a pilha ver Cap 12 Dados t cnicos Preste aten o polaridade correcta e ao introduzir as pilhas Volte a fechar a tampa das pilhas Retire as pilhas do aparelho caso este n o seja utili...

Page 42: ...conseguir uma pres s o adequada Importante Durante todo o processo de medi o n o deve mover se nem falar Enquanto a press o na bra adei ra diminui indicado o s mbolo do cora o e a diminui o da pressao...

Page 43: ...por componentes de precis o de alta qualidade N o deixe cair o apa relho Evite fortes vibra es Nunca abra o aparelho As repa ra es s podem ser efectuadas por t cnicos qualificados Limpe o aparelho ex...

Page 44: ...i o n o deve mover se ou falar 12 Dados t cnicos Processo de medi o Oscilom trico Margem de indica o 0 280 mmHg Zona de medi o S stole SYS 30 270 mmHg Di stole DIA 30 250 mmHg Pulsa o 40 199 pulsa es...

Page 45: ...qual tem como base a norma de produtos medicinais 93 42 EWG e possui a marca CE Entre outros o aparelho correspon de ao regulamento da directiva europeia aparelhos para medi o de tens o arterial n o...