Hardy Process Solutions HI 6020IT User Manual Download Page 10

HI 6020IT and HI 6020JB User Guide 

 

Page | 10 

 

 

Connecting Load Cell Sensors to a Hardy Summing Box 

The  diagrams  below  show  Hardy  Load  Sensors  with  C2  and  non-Hardy  Load  Cells  (4  wire  and  6  wire  are 
similar except 6 wire adds sense wiring) that do not have C2.  Wire the terminal connector that plugs into the 
front of the Weight Processor Modules by carefully following the Weigh Scale Input termination label. 

 

 

Non-C2 load sensor wiring

 

 
 

 

Hardy load sensor with C2 wiring 

 

WARNING: 

To prevent ignition of flammable or combustible atmospheres, disconnect power before 

servicing 

 

CAUTION: 

Observe precautions for electrostatic sensitive devices when handling the summing box 

terminal connections  

 

 

 

Strain Gauge

Load Cell

Load Cell

Shield

-Exc

+Exc

No Connect

-Sig

+Sig

Shield

+Sen

-Sen

No Connect

-Exc

+Exc

-Sig

+Sig

Factory Installed Jumpers

Hardy Weigh Scale
Input Connector

Load Cell

Shield

-Exc

+Exc

-Sig

+Sig

Shield

+Sen

-Sen

-Exc

+Exc

-Sig

+Sig

Factory Installed Jumpers

Hardy Weigh Scale
Input Connector

+C2

-C2

Strain Gauge

C2 Load Cell

+C2

-C2

Summary of Contents for HI 6020IT

Page 1: ...nd HI 6020JB US Canada Hazardous Locations User Guide Class I II III Division 1 Class I Zone 0 and Zone 20 Hardy Process Solutions Document Number 0596 0352 01 REV B CONTROLLED DOCUMENT CHANGES MAY NO...

Page 2: ...L WHO ARE FAMILIAR WITH THE CONSTRUCTION AND OPERATION OF ALL COMPONENTS IN THE SYSTEM AND THE POTENTIAL HAZARDS INVOLVED FAILURE TO OBSERVE THIS PRECAUTION COULD RESULT IN BODILY HARM AND OR PROPERTY...

Page 3: ...abilities 7 Hardy Process Toolbox 7 C2 and Calibration 7 INTEGRATED TECHNICIAN 7 Chapter 3 8 Installation of Intrinsically Safe HI 6020 IT and HI 6020JB Summing Boxes into Hazardous Classified Locatio...

Page 4: ...ox is a critical component in a weighing system that enables use of Hardy s core technologies C2 electronic calibration and Integrated Technician IT Each summing box distributes excitation voltage to...

Page 5: ...thout trim pots or X 2 with trim pots Y Y blank is a 5 hole summing box enclosure Y 6 is a 6 hole summing box enclosure enabling connection to a second summing box The HI 6020JB and HI 6020IT is provi...

Page 6: ...iew the installation section for additional information Trim Pot Number of Turns 11 for use with non C2 systems Trim Pot Impedance Range 0 10 Power Rating Non Hazardous Safe area Locations Class I and...

Page 7: ...the need for test weights A C2 weighing system consists of up to eight load cell sensors per HI 6020 IT or JB summing box C2 interconnect cable and a Hardy instrument controller or module with C2 Each...

Page 8: ...atus manufacturer s installation drawing must be followed when installing this equipment 2 Resistance between Intrinsically Safe Ground and earth ground must be lessthan 1 0 Ohm 3 Install Intrinsic Sa...

Page 9: ...t controller or module you will need a summing cardorsummingboxtoaggregatethesignalstoprovideasingularweight reading Whileaplatform benchor floorscalehasabuiltinsummingcardtoperformthisfunction loadce...

Page 10: ...n C2 load sensor wiring Hardy load sensor with C2 wiring WARNING To prevent ignition of flammable or combustible atmospheres disconnect power before servicing CAUTION Observe precautions for electrost...

Page 11: ...D ASSOCIATED APPARATUS FOR USE IN CLASS I II III DIVISION 1 HAZARDOUS CLASSIFIED LOCATIONS Edition 8 Revision Date 2015 04 28 CSA C22 2 NO 157 92 INTRINSICALLY SAFE AND NON INCENDIVE EQUIPMENT FOR USE...