2.3 Basic Waveform Output Example
Configure the generator to output a pulse waveform with 1.5MHz frequency, 500mVpp amplitude,
5mV
DC
DC offset, 200ns pulse width, 75ns leading edge time and 100ns trailing edge time.
1. Press
[CH1/2/3]
button on the front panel and select CH1.The corresponding area of CH1 in
the user interface is illuminated.
2. Press
[Wave] > Pulse
softkey to select the pulse wave function.
3. Press
Frequency
softkey to highlight it.Use the numeric keypad to input the frequency value
"1.5", and then select the desired unit "MHz" from the pop-up menu.
4. Press
Amplitude
softkey to highlight it.Use the numeric keypad to input the amplitude value
"500", and then select the desired unit "mVpp" in the pop-up menu.
5. Press
Offset
softkey to highlight it.Use the numeric keypad to input the offset value "5", and
then select the desired unit "mV" in the pop-up menu.
6. Press
Width
softkey to highlight it.Use the numeric keypad to input the width value "200", then
select the unit "nsec" from the menu that pops up.
7. Press
Edge Time
softkey to enter the edge time submenu.Press
Leading
softkey to highlight
it.Use the numeric keypad to enter the value "75", then select the unit "nsec" from the menu
that pops up.
8. Press
Edge Time
softkey to enter the edge time submenu.Press
Trailing
softkey to highlight
it.Use the numeric keypad to enter the value "100", then select the unit "nsec" from the menu
that pops up.
9. Press
[Output1]
button on the front panel to turn CH1 Output on.
10. At this point, the specified waveform is output from CH1 based on the current
configuration.Connect the CH1 output terminal to the oscilloscope and the waveform is as
shown in the figure below.
User Manual 22