13
• After confirmation, the selected param-
eter is displayed. The user can scroll
through the parameters by pressing
CFM.
In order to directly set another pa-
rameter, press SETUP again and en-
ter the code or scroll to it by pressing
CFM.
The following table lists the setup codes along with the
description of the specific setup items, their valid values
and whether password is required to view that item (“PW”
column):
Code
Valid Values
Default
PW
00 Factory ID
0 to 9999
0000
no
01 Process ID
0 to 9999
0000
no
02 Control
enable/disable
0: C.M. disabled
1: C.M. enabled
0
no
11 Relay 1 mode (M1)
0: disabled
1: ON-OFF high setpoint
2: ON-OFF low setpoint
3: Proportional, high setpoint
4:Proportional, low setpoint
0
no
12 Relay 1 setpoint (S1)
0.00 to 14.00 pH
-2000 to 2000 mV
8.00 pH
500 mV
no
13 Relay 1 hysteresis (H1)
0.00 to 14.00 pH
0 to 4000 mV
1 pH
50 mV
no
14 Relay 1 deviation (D1)
0 to 50 to 14.00 pH
25 to 4000 mV
1 pH
50 mV
no
21* Relay 2 mode (M2)
same as relay 1
0
no
22* Relay 2 setpoint (S2)
0.00 to 14.00 pH
-2000 to 2000 mV
6.00 pH
-500 mV
no
23* Relay 2 hysteresis (H2)
0.00 to 14.00 pH
0 to 4000 mV
1 pH
50 mV
no
24* Relay 2 deviation (D2)
0.50 to 14.00 pH
25 to 4000 mV
1 pH
50 mV
no
* Available only in models with two relays