• Press CFM to confirm the selected parameter that will stop
blinking on the primary display. The secondary display
shows the HI931002 or multimeter input value as lower
limit of the interval.
• Use the
or
to make the HI931002 or multimeter
output correspond with the meter’s value shown on the
secondary display (e.g. 4).
• Wait for approximately 30 seconds (until the reading of
the calibrator is stable).
• Press CFM to enter. The meter will switch to the second
calibration point. Repeat the above procedure.
• After the desired readings are obtained, press CFM and
the meter will skip back to normal operating mode.