31
CALIBRA
TION
Press
CAL
from any measure mode (DO, BOD, OUR
or SOUR). The calibration menu will be displayed.
Press
Pressure
functional key to select the pressure calibration. The pressure calibration screen will
be displayed.
Using the / keys, enter the true local barometric pressure reading from the reference barometer.
DO NOT
use the pressure reported by the weather bureau. Weather bureaus correct pressures to sea
level.
When the reading is stable within range of the entered barometric pressure the
CFM
functional key
is displayed.
Press
CFM
to confirm the calibration.
The meter will return to the measurement screen and will store the calibration data.
”CAL DUE” will be displayed and a probe calibration must follow.
To clear the pressure calibration and restore the factory one, press
CAL
from any measure mode (DO,
BOD, OUR or SOUR).
The calibration menu will be displayed. Press the
Pressure
functional key and then
Clear
.
The user calibration will be removed and the factory calibration restored.
Press
ESC
key at any time to exit the calibration without changing values.
9.3. TEMPERATURE CALIBRATION
The
HI764113
has been factory calibrated for temperature. Dissolved oxygen values are based upon
temperature compensation so accurate temperature measurements are required. Should a thermistor
go out of tolerance, your measurement will be compromised. An additional user temperature
calibration of the
HI764113
optical dissolved oxygen probe may be performed if desired.