58
•
SOUR temperature correction (1=SOUR@20°C, 0=SOUR not corrected) (1 char)
•
BOD DATA log data
:
•
Sample type (1 = sample, 0 = seed) (1 char)
•
Bottle ID (4 chars)
•
DO value, including sign and decimal point [mg/L] (8 chars)
•
Bottle volume, including sign and decimal point [ml] (6 chars)
•
Sample volume, including sign and decimal point [ml] (6 chars)
•
Seed volume, including sign and decimal point [ml] (6 chars)
•
Salinity value [g/L] (3 chars)
•
Pressure value in mmHg, including sign and decimal point (11 chars)
•
Temperature value, including sign and decimal point (8 chars)
•
Log time: yy mm dd hh mm ss (12 chars)
Notes: “Err8” is sent if the instrument is not in measurement mode.
“Err6” is sent if the requested range is not available.
“Err4” is sent if the requested set parameter is not available.
“Err3” is sent if the Log on demand is empty.
“Err9” is sent if the battery power is less than 30%.
Invalid commands will be ignored.
PC INTERF
ACE
www.tools.in.th
www.neonics.co.th