56
NSLx
Requests the number of logged samples (4 chars)
The command parameter (x ‑ 1 char)
•
D
‑ the request is for DO
•
B
‑ the request is for BOD
•
O
‑ the request is for OUR
•
S
‑ the request is for SOUR
•
I
‑ the request is for BOD initial values
LODDxxx
Requests the xxx
th
DO logged data
LODBxxx
Requests the xxx
th
BOD logged data
LODOxxx
Requests the xxx
th
OUR logged data
LODSxxx
Requests the xxx
th
SOUR logged data
LODIxxx
Requests the xxx
th
initial BOD value logged data
LODDALL
Requests all the DO log on demand
LODBALL
Requests all the BOD log on demand
LODOALL
Requests all the OUR log on demand
LODSALL
Requests all the SOUR log on demand
LODIALL
Requests all the initial BOD values log
The answer string for each record contains:
•
The logged mode (2 chars)
•
20 ‑ DO range
•
21 ‑ BOD range
•
22 ‑ OUR range
•
23 ‑ SOUR range
•
24 ‑ BOD initial values
•
DO log data
:
•
Measurement unit (0 = %, 1 = mg/L) (1 char)
•
DO value including sign and decimal point (8 chars)
•
Salinity value [g/L] (3 chars)
•
Pressure value in mmHg, including sign and decimal point (11 chars)
•
Temperature value including sign and decimal point (8 chars)
•
BOD log data
:
•
Sample type (1 = sample, 0 = seed) (1 char)
•
Seed corrected (1 = corrected, 0 = not corrected) (1 char)
•
Bottle ID (4 chars)
•
BOD value including sign and decimal point [mg/L] (8 chars)
•
Bottle volume including sign and decimal point [ml] (6 chars)
PC INTERF
ACE
www.tools.in.th
www.neonics.co.th