PAR
Requests the setup parameters setting.
The answer string contains:
•
backlight value (1 char)
•
contrast value (2 chars)
•
instrument ID (4 chars)
•
calibration alarm timeout (2 chars)
•
Setup information (2 chars): 8 bit hexadecimal encoding:
•
0x01:
‑ beep is on
•
0x04:
‑ Celsius / Fahrenheit display (°C if the bit is set)
•
0x10:
‑ manual pressure (1 if activated, 0 otherwise)
•
Auto Light Off time (3 chars)
•
Auto Power Off time (3 chars)
•
Salinity value (3 chars)
•
Pressure unit (1 char): 0 ‑ mmHg, 1 ‑ inHg, 2 ‑ atm, 3 ‑ mbar, 4 ‑ psi, 5 ‑ kPa
•
BOD
configuration values
•
sample minimum delta DO, including sign and decimal point (6 chars
)
•
sample minimum end DO, including sign and decimal point (6 chars)
•
seed minimum delta DO, including sign and decimal point (6 chars)
•
seed minimum end DO, including sign and decimal point (6 chars)
•
OUR
configuration values
•
minimum time in seconds (4 chars)
•
maximum time in seconds (4 chars)
•
minimum start DO including sign and decimal point (6 chars)
•
minimum end DO including sign and decimal point (6 chars)
•
total volume including sign and decimal point (6 chars)
•
sample volume including sign and decimal point (6 chars)
•
SOUR
configuration values
•
minimum time in seconds (4 chars)
•
maximum time in seconds (4 chars)
•
minimum start DO including sign and decimal point (6 chars)
•
minimum end DO including sign and decimal point (6 chars)
•
total volume including sign and decimal point (6 chars)
•
sample volume including sign and decimal point (6 chars)
•
solids weight including sign and decimal point (6 chars)
•
SOUR temperature correction (1=enabled, 0=disabled) (1 char)
•
The short name of the selected language (3 chars)
PC INTERF
ACE