36
Manual Pressure
Highlight
Manual pressure.
Press the displayed functional key to enable/disable the feature.
If enabled, the pressure can be entered by the user, while in measurement screen, using
/ keys.
Pressure Unit
Highlight
Pressure unit
.
Press
Modify
.
Autodelete BOD Start Data
Highlight
Autodelete BOD start data.
Press the displayed functional key to enable/disable the feature.
If enabled the BOD initial data record used in BOD result evaluation is deleted automatically after the
BOD result has been saved into the instrument’s memory (
LOG
key pressed).
If disabled, the user has to delete BOD initial data records that were used in BOD result evaluation,
entering
View initial BOD data
mode.
SETUP
www.tools.in.th