Stand-Alone Tester (Up to Three-Point Calibration)

1.  Rinse the electrode tip with purified water and blot dry. Then rinse with the buffer being used for calibration.

2.  Press and hold ON/OFF button until “

CAL

 

MODE

” is displayed.

3.  When “

7.01 

USE

” or “

6.86 

USE

” is displayed with “CAL” tag blinking, place the tip of the electrode in the 

correct buffer.

CAL

MODE

   

7.01

USE

4.  When the buffer has been recognized, “

REC

“ is displayed. 

WAIT

” is displayed with the stability indicator (

#

) blinking until the reading is stable.

Wait until the measurement is stored and the stability indicator disappears.

 

7.01

REC

   

#

7..01

WAIT

5.  To save a one-point calibration and return to measurement mode, press the ON/OFF button.

CAL

 

SAVE

” is displayed briefly.

With 2P option selected

6.  Place the tip of the electrode in the second buffer rinse beaker, then in the second calibration buffer. Wait until the 

measurement is stored and the stability indicator disappears. 

7.  After the second point has been stored, “

CAL

 

SAVE

” is displayed briefly and the tester will return to measurement 

mode automatically.

With 3P option selected

 

6.  Place the tip of the electrode in the second buffer rinse beaker, then in the second calibration buffer. Wait until the 

measurement is stored and the stability indicator disappears. 

7.  Place the tip of the electrode in the third buffer rinse beaker, then in the third calibration buffer. Wait until the 

measurement is stored and the stability indicator disappears. 

8.  After the third point has been stored, “

CAL

 

SAVE

” is displayed briefly and the tester will return to measurement 

mode automatically.

With Hanna Lab App (Up to Four-Point Calibration)

Connect the tester to the Hanna Lab App and follow the calibration procedure.

See the App Help section for the calibration procedure.

Summary of Contents for HI9810342

Page 1: ...INSTRUCTION MANUAL HI9810342 Wireless pH Tester for Milk with built in specialized electrode Foodcare...

Page 2: ...a Hanna Instruments product Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument a...

Page 3: ...ffer Set 6 pH Resolution 6 Bluetooth Wireless Mode 6 Bluetooth Pairing 7 Date Time 7 7 Bluetooth 7 8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 7 St...

Page 4: ...pH 0 01 or 0 1 pH mV 0 1 or 1 mV Temperature 0 1 C 0 1 F Accuracy pH 0 05 pH Temperature 0 5 C 0 9 F Calibration Up to three points or four points Automatic buffer recognition with Standard buffers H...

Page 5: ...342 features a non porous glass body that is easy to clean and disinfect The conical tip allows for easy penetration into semisolids ideal for milk and milk products like yogurt Specialized low temper...

Page 6: ...d battery type Dispose of old battery in accordance with local regulations 6 SETUP The Setup button is located inside the battery compartment After Setup configuration replace the cover Setup Menu Nav...

Page 7: ...able mode it will appear in the list of Available Devices Within the application tap Connect to pair the tester with the device All readings are transmitted directly to the application Tag a Measureme...

Page 8: ...e in the second buffer rinse beaker then in the second calibration buffer Wait until the measurement is stored and the stability indicator disappears 7 After the second point has been stored CAL SAVE...

Page 9: ...not in use add a few drops of storage solution to storage cap If storage solution is not available pH 4 01 or pH 7 01 buffer can be used Note Do not store the electrode in distilled or deionized water...

Page 10: ...d for the environment in which it is used Any variation introduced by the user to the supplied equipment may degrade the tester s performance For your and the tester s safety do not use or store it in...

Page 11: ...12208A 04 This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must a...

Page 12: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN9810342 Printed in ROMANIA...

Reviews: