Hanna Instruments HI9810342 Instruction Manual Download Page 4

4

Pr

EL

imi

NA

ry Ex

Ami

NA

Ti

ON

1.  PRELIMINARY EXAMINATION

Remove the tester and accessories from packaging and examine them carefully. For further assistance, please contact your 

local Hanna Instruments office or email us at [email protected].
Each 

HI9810342

 is supplied with a starter kit consisting of:

•  pH 4.01 buffer solution, 20 mL sachet (2 pcs.)

•  pH 7.01 buffer solution, 20 mL sachet (2 pcs.)

•  Cleaning solution for milk deposits, 20 mL sachet (2 pcs.)

•  Electrode storage solution, 13 mL dropper bottle (1 pc.)

•  3V Lithium battery – CR2032

•  Instrument quality certificate and Instruction manual

Note:

 Save all packing material until you are sure that the tester works correctly. Any damaged or defective item must be 

returned in its original packing material with the supplied accessories.

2.  SPECIFICATIONS

Range

pH

0.00 to 12.00 pH

mV *

pH/mV conversion

Temperature ** 0.0 to 60.0 °C (32.0 to 140.0 °F)

Resolution

pH

0.01 or 0.1 pH

mV *

0.1 or 1 mV

Temperature

0.1 °C; 0.1 °F

Accuracy 

pH

±0.05 pH 

Temperature

±0.5 °C; ±0.9 °F

Calibration

Up to three points or four points *

Automatic buffer recognition with Standard buffers

Hanna (pH 1.68 *, 4.01, 7.01, 10.01) or NIST (pH 1.68 *, 4.01, 6.86, 9.18)

Temperature compensation Automatic (ATC) or Manual (MTC) *

Electrode

Body material 

Glass

Glass 

Low Temperature (LT)

Junction Open

Reference cell 

Double, Ag/AgCl

Electrolyte Gel

Tip / Shape 

Conic, Ø 12 x 12 mm (Ø 0.47 x 0.47”)

Outer diameter 

12 mm (0.5”)

Length 

85 mm (3.3”)

Battery type 

3V Lithium – CR2032

Battery life

Approximately 1000 hours (500 hours with Bluetooth enabled)

Environment

0 to 50 °C (32 to 122 °F)

Casing

IP65 ingress protection

Dimensions / Weight

51 x 160 x 21 mm (2.0 x 6.3 x 0.8“) / 50 g (1.8 oz.)

* Available with Hanna Lab App

** Measuring outside the recommended operating temperature range may damage the gel electrolyte and void product warranty.

Note:

 The tester can display measurements from 

2.00 to 16.00 pH. Measurements outside of the pH range will flash. In this 

case assess the integrity of the tester and the type of measured sample.

Summary of Contents for HI9810342

Page 1: ...INSTRUCTION MANUAL HI9810342 Wireless pH Tester for Milk with built in specialized electrode Foodcare...

Page 2: ...a Hanna Instruments product Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument a...

Page 3: ...ffer Set 6 pH Resolution 6 Bluetooth Wireless Mode 6 Bluetooth Pairing 7 Date Time 7 7 Bluetooth 7 8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 7 St...

Page 4: ...pH 0 01 or 0 1 pH mV 0 1 or 1 mV Temperature 0 1 C 0 1 F Accuracy pH 0 05 pH Temperature 0 5 C 0 9 F Calibration Up to three points or four points Automatic buffer recognition with Standard buffers H...

Page 5: ...342 features a non porous glass body that is easy to clean and disinfect The conical tip allows for easy penetration into semisolids ideal for milk and milk products like yogurt Specialized low temper...

Page 6: ...d battery type Dispose of old battery in accordance with local regulations 6 SETUP The Setup button is located inside the battery compartment After Setup configuration replace the cover Setup Menu Nav...

Page 7: ...able mode it will appear in the list of Available Devices Within the application tap Connect to pair the tester with the device All readings are transmitted directly to the application Tag a Measureme...

Page 8: ...e in the second buffer rinse beaker then in the second calibration buffer Wait until the measurement is stored and the stability indicator disappears 7 After the second point has been stored CAL SAVE...

Page 9: ...not in use add a few drops of storage solution to storage cap If storage solution is not available pH 4 01 or pH 7 01 buffer can be used Note Do not store the electrode in distilled or deionized water...

Page 10: ...d for the environment in which it is used Any variation introduced by the user to the supplied equipment may degrade the tester s performance For your and the tester s safety do not use or store it in...

Page 11: ...12208A 04 This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must a...

Page 12: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN9810342 Printed in ROMANIA...

Reviews: