5.  GENERAL OPERATIONS

Turning the Tester On & Off

•  Turn the tester face down. Rotate the battery cover on the back of the tester counterclockwise and set it aside. Remove 

the battery insulation film. 

•  Press the ON/OFF button to turn the tester on or off. Initialization screen displays all LCD segments, followed by the 

battery percentage. The tester enters measurement mode. Before the tester turns off, “

OFF

 

PWR

” is displayed briefly.

•  Press and hold the ON/OFF button to turn the tester off when connected to Bluetooth.

Battery Replacement

1.  Turn off the tester. Turn the tester face down and rotate the battery cover counterclockwise. 

2.  Set the battery cover aside. Press the metallic pin to push old battery out.

3.  Place the new battery with positive (+) sign facing out.

4.  Reset the date and time in setup, or connect to the Hanna Lab App to update it automatically.

5.  Align the mark on the cover with the open lock icon ( ) on the case. Rotate the cover 

clockwise until the mark on the cover aligns with the closed lock icon ( ).

Note:

 Only use specified battery type. Dispose of old battery in accordance with local regulations.

6.  SETUP

The Setup button is located inside the battery compartment. After Setup configuration, replace the cover.

Setup Menu Navigation

•  Press the Setup button to enter Setup mode and navigate menu items.

•  To exit Setup mode, press the Setup button after “

SEt

 

TIME

” option is displayed.

•  Press the ON/OFF button to configure menu item options.

Temperature Unit 

Option:

 °C or °F

Press ON/OFF button to select desired 

temperature unit.

°C

T E M P

 

°F

TEMP

Auto-Off Interval

Option:

 8, 60 min., or “

----

” (disabled)

Press ON/OFF button to select desired interval.

To save battery life, after the selected auto-off

8

AOFF

 

60

AOFF

 

----

AOFF

interval has elapsed, the tester will automatically turn off. 

Calibration Points

Option:

 2P or 3P

Press ON/OFF button to select between up to 

two- or up to three-point calibration.

2P

CAL

 

3P

CAL

Buffer Set

Option:

 7.01 pH (Hanna) or 6.86 pH (NIST)

Press ON/OFF button to select the calibration 

buffer set (Hanna or NIST).

7.01

BUFF

 

6.86

BUFF

pH Resolution

Option:

 0.01 pH or 0.1 pH

Press the ON/OFF button to select resolution.

0.01

RES

 

0.1

RES

Bluetooth Wireless Mode

Option:

 On, PAIr, or OFF

Press ON/OFF button to select Bluetooth 

option at start-up.

&

On

BLU

 

&

PAIr

BLU

 

&

OFF

BLU

Metallic pin

Summary of Contents for HI9810342

Page 1: ...INSTRUCTION MANUAL HI9810342 Wireless pH Tester for Milk with built in specialized electrode Foodcare...

Page 2: ...a Hanna Instruments product Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument a...

Page 3: ...ffer Set 6 pH Resolution 6 Bluetooth Wireless Mode 6 Bluetooth Pairing 7 Date Time 7 7 Bluetooth 7 8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 7 St...

Page 4: ...pH 0 01 or 0 1 pH mV 0 1 or 1 mV Temperature 0 1 C 0 1 F Accuracy pH 0 05 pH Temperature 0 5 C 0 9 F Calibration Up to three points or four points Automatic buffer recognition with Standard buffers H...

Page 5: ...342 features a non porous glass body that is easy to clean and disinfect The conical tip allows for easy penetration into semisolids ideal for milk and milk products like yogurt Specialized low temper...

Page 6: ...d battery type Dispose of old battery in accordance with local regulations 6 SETUP The Setup button is located inside the battery compartment After Setup configuration replace the cover Setup Menu Nav...

Page 7: ...able mode it will appear in the list of Available Devices Within the application tap Connect to pair the tester with the device All readings are transmitted directly to the application Tag a Measureme...

Page 8: ...e in the second buffer rinse beaker then in the second calibration buffer Wait until the measurement is stored and the stability indicator disappears 7 After the second point has been stored CAL SAVE...

Page 9: ...not in use add a few drops of storage solution to storage cap If storage solution is not available pH 4 01 or pH 7 01 buffer can be used Note Do not store the electrode in distilled or deionized water...

Page 10: ...d for the environment in which it is used Any variation introduced by the user to the supplied equipment may degrade the tester s performance For your and the tester s safety do not use or store it in...

Page 11: ...12208A 04 This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must a...

Page 12: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN9810342 Printed in ROMANIA...

Reviews: