background image

Main Features 

4

TDS

 is a calculated value based on the conductivity of the solution (TDS = factor x EC

25

). A TDS factor is a 

conversion factor used to change an EC measurement to a ppm measurement.

Sal psu

: The practical salinity of seawater relates the ratio of electrical conductivity of a normal seawater sample 

at 15 ºC and 1 atmosphere to a potassium chloride solution (KCl) with a mass of 32.4356 g/Kg water at the 
same temperature and pressure. Under these conditions the ratio is equal to 1 and S=35. The practical salinity 
scale may be applied to values 0 through 42.00 psu at temperatures between 0 to 35 ºC.

Sal ppt:

 measurements expressed in 

ppt

 are based on the Natural Seawater Scale that extends from 0.00 to 

80.00 g/L and covers 10 to 31 °C temperature range. It determines the salinity based upon a conductivity ratio 
of sample to standard seawater at 15 °C and an approximate salinity value of 35 in seawater.

Sal %: 

in this scale 100% salinity is equivalent to roughly 10% solids.

The conductivity probes can be installed directly in-line, immersed in a tank, or in a flow cell, and are suited 
for continuous measurement of conductivity and associated parameters required in applications such as water 
treatment, drinking water, feed water condensate, or other clean water applications as well as sea water and 
surface-water monitoring.
The probe is available with an 

integral connector

 that permits connections with no twisted cables as cables 

are connected after installation. 

Patch cables

 may be purchased separately to connect between the probe and controller up to 50 meters (164 ft). 

4.  MAIN FEATURES

•  Rugged, chemically-resistant PVDF body
• 

3/4

” NPT external thread for mounting

•  6 bar (87 psi) maximum pressure at 25 °C (77 °F)
•  Digital probe stores model, firmware, serial number, and calibration information

5.  SERIES CONFIGURATION

HI7630

– y

8

z z

y

2

   Two-electrode cell conductivity, AISI 316 stainless steel 

cell constant k ≈ 0.1/cm

4

   Four-ring conductivity, platinum on glass  

cell constant k ≈ 1.0/cm

8

Smart probe

 with RS485 connection

zz

00

 supplied with DIN connector (without cable)*

05

10

15

25

50

 fixed cable length (in meters)

See Accessories section for patch cable ordering codes.

Summary of Contents for HI7630-28

Page 1: ...4805 0 to 30 mS cm T 0 to 50 C p max 6 bar SN XXXXXXXXXXXX HI7630 2800 Conductivity and Temperature Industrial Smart Probes Please scan the QR code or follow the link to download the HI510 controller...

Page 2: ...intenance 14 13 Accessories 15 14 Abbreviations 16 Certification 16 Recommendations for Users 16 Warranty 16 Dear Customer Thank you for choosing a Hanna Instruments product Please read this instructi...

Page 3: ...ons detailed in this manual 3 GENERAL DESCRIPTION INTENDED USE HI7630 28 and HI7630 48 are Conductivity and Temperature probes designed to be paired with the HI510 Universal Process Controller The pro...

Page 4: ...conductivity probes can be installed directly in line immersed in a tank or in a flow cell and are suited for continuous measurement of conductivity and associated parameters required in applications...

Page 5: ...h standard solution Process Single point Temperature Range 0 0 to 50 0 C 32 0 to 122 0 F Accuracy 0 5 C 1 0 F Calibration Single point offset controller setting Temperature compensation mode Linear Na...

Page 6: ...ion Salinity Single point with standard solution Resistivity Single point Process Single point Temperature Range 0 0 to 100 0 C 32 0 to 212 0 F Accuracy 0 5 C 1 0 F Calibration Single point offset con...

Page 7: ...189 mm 7 4 Four Ring Probes HI7630 48zz 28 mm 1 1 72 mm 2 8 50 mm 2 12 mm 0 5 3 4 NPT 216 mm 8 5 HI7630 4800 216 mm 8 5 Align the pins and key then push the plug into the socket Rotate the collar to...

Page 8: ...essibility for maintenance when selecting placement Typical installation examples with mounting accessories Accessories are sold separately In line mounting and flow cell installation require that the...

Page 9: ...PT 2 Vent hole 3 O ring 4 Probe Tank immersion with submersible electrode holder I II III 1 2 4 3 5 6 7 9 8 1 Probe cable 2 Cable gland 3 Electrode holder cap 4 Adjustable immersion level 5 Electrode...

Page 10: ...3 and cable gland on top 2 The cable is shielded inside the holder to prevent any damage to the insulation F Tighten the cable gland 2 step III and slide the cap 3 back onto the electrode holder 5 Fl...

Page 11: ...n the board 4 Position excess cable through the cable gland before tightening the nut 5 Remove the ground screw and hardware located below the PROBE connector and attach ground lead g COMM B A IN1 IN2...

Page 12: ...oss contamination when a two point calibration is required use two beakers one for rinsing the probe and the other for calibration Go to Probe Settings in Channel Setup and set Meas Mode to EC Procedu...

Page 13: ...ed in conductivity EC Meas Mode The user can subsequently change Meas Mode to RES to measure in resistivity units M cm Ultrapure water self ionizes into H and OH ions and has a conductivity of 0 055 S...

Page 14: ...ccurate readings Remove and inspect the probe during scheduled service intervals Dry clean the sensor with a soft bristle brush to loosen any debris For a more thorough cleaning Use a cloth and warm w...

Page 15: ...1 2 PVC protective end cap inside height 68 mm 2 6 HI60503 PVDF immersion electrode holder HI60503 2 PVDF protective end cap inside height 68 mm 2 6 HI605011 PVC mounting flange HI60542 In line electr...

Page 16: ...oduced by the user to the supplied equipment may degrade its performance Do not use or store the product in hazardous environments WARRANTY The probes are warrantied for six months against defects in...