Preliminary Examination

1.  PRELIMINARY EXAMINATION

Remove the probe from the packing and examine it carefully. 
Each probe is supplied with quick reference guide and probe quality certificate.

HI7630-28

 two-ring probes are supplied with a protective cap on. Remove the cap prior to use.

HI7630-48

 four-ring probes are fitted with a plastic support bung located at the end of the probe. Remove 

the bung prior to use.

Note

: Save all packing material until you are sure that the probe works correctly. Any damaged or defective item 

must be returned in its original packing material with the supplied accessories.

2.  SAFETY MEASURES

General Safety & Installation Recommendations

•  Electrical connection, installation, operation, and maintenance must be carried out by specialized 

personnel only. 
The specialized personnel must read, understand, and adhere to the instructions in this manual. 

•  Before connecting the probe to the process controller, disconnect the controller from the power line.
•  User serviceable connections are all accessible inside controller’s enclosure.
•  Before powering the controller, verify the probe wiring has been done following the instructions 

detailed in this manual.

3.  GENERAL DESCRIPTION & INTENDED USE

HI7630-28

 and 

HI7630-48

 are Conductivity and Temperature probes designed to be paired with the 

HI510

 Universal Process Controller.

The probes are recommended for a wide range of water applications, from pure and ultrapure water (two-ring 
series i.e. 

HI7630-28

) to industrial process water (four-ring series i.e. 

HI7630-48

 ).

The two-ring probes can be calibrated using a standard with a value close to the measurement value whereas the 
four-ring probes provide stable measurements over a wide measure range and do not require frequent calibrations.
An integral 

temperature sensor

 measures the process temperature and adjusts the measured conductivity to 

a reference temperature by applying specialized compensation standards:

• 

Linear

: appropriate when it is assumed that the temperature coefficient of variation has the same value 

for all measurement temperatures. 

• 

Standard

: appropriate for high-purity water measurements and documented in ASTM Standard D5391-14. 

This setting should be used for resistivity measurements. 

• 

Natural

: appropriate for natural ground, well, or surface water (or water with similar composition) in 

accordance with ISO7888 standard.

The result is reliable electrolytic conductivity (EC), TDS (Total Dissolved Solids), resistivity, or Seawater Salinity 
in percent, psu, or ppt units.

Summary of Contents for HI7630-28

Page 1: ...4805 0 to 30 mS cm T 0 to 50 C p max 6 bar SN XXXXXXXXXXXX HI7630 2800 Conductivity and Temperature Industrial Smart Probes Please scan the QR code or follow the link to download the HI510 controller...

Page 2: ...intenance 14 13 Accessories 15 14 Abbreviations 16 Certification 16 Recommendations for Users 16 Warranty 16 Dear Customer Thank you for choosing a Hanna Instruments product Please read this instructi...

Page 3: ...ons detailed in this manual 3 GENERAL DESCRIPTION INTENDED USE HI7630 28 and HI7630 48 are Conductivity and Temperature probes designed to be paired with the HI510 Universal Process Controller The pro...

Page 4: ...conductivity probes can be installed directly in line immersed in a tank or in a flow cell and are suited for continuous measurement of conductivity and associated parameters required in applications...

Page 5: ...h standard solution Process Single point Temperature Range 0 0 to 50 0 C 32 0 to 122 0 F Accuracy 0 5 C 1 0 F Calibration Single point offset controller setting Temperature compensation mode Linear Na...

Page 6: ...ion Salinity Single point with standard solution Resistivity Single point Process Single point Temperature Range 0 0 to 100 0 C 32 0 to 212 0 F Accuracy 0 5 C 1 0 F Calibration Single point offset con...

Page 7: ...189 mm 7 4 Four Ring Probes HI7630 48zz 28 mm 1 1 72 mm 2 8 50 mm 2 12 mm 0 5 3 4 NPT 216 mm 8 5 HI7630 4800 216 mm 8 5 Align the pins and key then push the plug into the socket Rotate the collar to...

Page 8: ...essibility for maintenance when selecting placement Typical installation examples with mounting accessories Accessories are sold separately In line mounting and flow cell installation require that the...

Page 9: ...PT 2 Vent hole 3 O ring 4 Probe Tank immersion with submersible electrode holder I II III 1 2 4 3 5 6 7 9 8 1 Probe cable 2 Cable gland 3 Electrode holder cap 4 Adjustable immersion level 5 Electrode...

Page 10: ...3 and cable gland on top 2 The cable is shielded inside the holder to prevent any damage to the insulation F Tighten the cable gland 2 step III and slide the cap 3 back onto the electrode holder 5 Fl...

Page 11: ...n the board 4 Position excess cable through the cable gland before tightening the nut 5 Remove the ground screw and hardware located below the PROBE connector and attach ground lead g COMM B A IN1 IN2...

Page 12: ...oss contamination when a two point calibration is required use two beakers one for rinsing the probe and the other for calibration Go to Probe Settings in Channel Setup and set Meas Mode to EC Procedu...

Page 13: ...ed in conductivity EC Meas Mode The user can subsequently change Meas Mode to RES to measure in resistivity units M cm Ultrapure water self ionizes into H and OH ions and has a conductivity of 0 055 S...

Page 14: ...ccurate readings Remove and inspect the probe during scheduled service intervals Dry clean the sensor with a soft bristle brush to loosen any debris For a more thorough cleaning Use a cloth and warm w...