Specifications 

6

HI7630-48

zz  Four-Ring Probe 

Range

EC 

0.0 µS/cm to 999.9 mS/cm*

TDS 

0.0 mg/L to 400.0 g/L (TDS factor 0.5)*

RES 

1.00 Ω

cm to 9.99 MΩ

cm*

Seawater Salinity 

400.0 %NaCl, 42 psu, 80 ppt

Resolution**

EC 

0.1 µS/cm; 0.001, 0.01, 0.1 mS/cm

RES 

0.1, 1 Ω

cm; 0.01, 0.1, 1 kΩ

cm, 0.01MΩ

cm 

Accuracy

EC 

 ± 2 % of reading or ±1 µS/cm, whichever is greater

TDS 

± 2 % of reading or ±0.5 mg/L, whichever is greater

Calibration

Standard 

Automatic, two-point with standard solution

Salinity 

Single point with standard solution

Resistivity 

Single point

Process 

Single point

Temperature

Range 

0.0 to 100.0 °C (32.0 to 212.0 °F)

Accuracy 

± 0.5 °C; ± 1.0 °F

Calibration 

Single point offset (controller setting)

Temperature 
compensation mode

Linear, Natural, Standard, None

Temperature source

Automatic (from probe)
Manual

Body

PVDF

Sensor

Type 

Four-ring, platinum 

Insulator Glass
Diameter 

12 mm (0.5”)

Insertion length 

50 mm (2”)

Maximum pressure

6 bar (87 psi)

Threaded connection

3/4

” NPT external thread for mounting

Protection rating

IP68

Cable length

zz

 , see Series Configuration

*

  Absolute values (not temperature compensated)

**

 Automatic ranging

Summary of Contents for HI7630-28

Page 1: ...4805 0 to 30 mS cm T 0 to 50 C p max 6 bar SN XXXXXXXXXXXX HI7630 2800 Conductivity and Temperature Industrial Smart Probes Please scan the QR code or follow the link to download the HI510 controller...

Page 2: ...intenance 14 13 Accessories 15 14 Abbreviations 16 Certification 16 Recommendations for Users 16 Warranty 16 Dear Customer Thank you for choosing a Hanna Instruments product Please read this instructi...

Page 3: ...ons detailed in this manual 3 GENERAL DESCRIPTION INTENDED USE HI7630 28 and HI7630 48 are Conductivity and Temperature probes designed to be paired with the HI510 Universal Process Controller The pro...

Page 4: ...conductivity probes can be installed directly in line immersed in a tank or in a flow cell and are suited for continuous measurement of conductivity and associated parameters required in applications...

Page 5: ...h standard solution Process Single point Temperature Range 0 0 to 50 0 C 32 0 to 122 0 F Accuracy 0 5 C 1 0 F Calibration Single point offset controller setting Temperature compensation mode Linear Na...

Page 6: ...ion Salinity Single point with standard solution Resistivity Single point Process Single point Temperature Range 0 0 to 100 0 C 32 0 to 212 0 F Accuracy 0 5 C 1 0 F Calibration Single point offset con...

Page 7: ...189 mm 7 4 Four Ring Probes HI7630 48zz 28 mm 1 1 72 mm 2 8 50 mm 2 12 mm 0 5 3 4 NPT 216 mm 8 5 HI7630 4800 216 mm 8 5 Align the pins and key then push the plug into the socket Rotate the collar to...

Page 8: ...essibility for maintenance when selecting placement Typical installation examples with mounting accessories Accessories are sold separately In line mounting and flow cell installation require that the...

Page 9: ...PT 2 Vent hole 3 O ring 4 Probe Tank immersion with submersible electrode holder I II III 1 2 4 3 5 6 7 9 8 1 Probe cable 2 Cable gland 3 Electrode holder cap 4 Adjustable immersion level 5 Electrode...

Page 10: ...3 and cable gland on top 2 The cable is shielded inside the holder to prevent any damage to the insulation F Tighten the cable gland 2 step III and slide the cap 3 back onto the electrode holder 5 Fl...

Page 11: ...n the board 4 Position excess cable through the cable gland before tightening the nut 5 Remove the ground screw and hardware located below the PROBE connector and attach ground lead g COMM B A IN1 IN2...

Page 12: ...oss contamination when a two point calibration is required use two beakers one for rinsing the probe and the other for calibration Go to Probe Settings in Channel Setup and set Meas Mode to EC Procedu...

Page 13: ...ed in conductivity EC Meas Mode The user can subsequently change Meas Mode to RES to measure in resistivity units M cm Ultrapure water self ionizes into H and OH ions and has a conductivity of 0 055 S...

Page 14: ...ccurate readings Remove and inspect the probe during scheduled service intervals Dry clean the sensor with a soft bristle brush to loosen any debris For a more thorough cleaning Use a cloth and warm w...

Page 15: ...1 2 PVC protective end cap inside height 68 mm 2 6 HI60503 PVDF immersion electrode holder HI60503 2 PVDF protective end cap inside height 68 mm 2 6 HI605011 PVC mounting flange HI60542 In line electr...

Page 16: ...oduced by the user to the supplied equipment may degrade its performance Do not use or store the product in hazardous environments WARRANTY The probes are warrantied for six months against defects in...

Reviews: