41
Holding Registers (4xxxx Addresses)
Address Register Function
Size
Min.
Value
Max.
Value
Default
Value Details
40196
Reset Controller status
uint32_t
bytes
1‑0
b0
= R/W Controller setup updated
b1
= R/W Chan1 calibration updated
b2
= R/W Chan2 calibration updated
b3
= RO‑ Chan1 probe update parameters
b4
= R/W ‑ Chan1 probe parameters updated
b5
= Cleaning active
b6
= Local edit mode active
b7
= Manual mode active
b8
= Chan1 calibration mode active
b9
= Remote Control Mode active
b10
= Remote Control Edit mode active
b11
= Controller in Hold mode
b12
= Controller in startup mode
b13
= Manual Hold active
b14
= External Hold active
40200
Math Channel Function
uint8_t
0
4
0
0
= Disabled
1
= Difference
2
= Ratio
3
= Passage
4
= Rejection
40201
Math Channel Alarm0_HighValue
float
bytes
1‑0
40202
bytes
3‑2
40203
Math Channel Alarm0_LowValue
float
bytes
1‑0
40204
bytes
3‑2
40205
Math Channel Alarm1_HighValue
float
bytes
1‑0
40206
bytes
3‑2
40207
Math Channel Alarm1_LowValue
float
bytes
1‑0
40208
bytes
3‑2
40209
Math Channel Alarm2_HighValue
float
bytes
1‑0
40210
bytes
3‑2
40211
Math Channel Alarm2_LowValue
float
bytes
1‑0
40212
bytes
3‑2
40213
Math Channel Alarm3_HighValue
float
bytes
1‑0
40214
bytes
3‑2
40215
Math Channel Alarm3_LowValue
float
bytes
1‑0
40216
bytes
3‑2
40217
Controller Clock_Year
uint8_t
20
99
22
40218
Controller Clock_Month
uint8_t
1
12
1
40219
Controller Clock_Day
uint8_t
1
31
1
40220
Controller Clock_WeekDay
uint8_t
0
6
1
40221
Controller Clock_Hour
uint8_t
0
23
12
40222
Controller Clock_Minute
uint8_t
0
59
0
40223
Controller Clock_Second
uint8_t
0
59
0